background image

 

Operation and Instruction Manual 

 

REV B  

 

 
 
 
 

Koehler Instrument Company, Inc. 

1595 Sycamore Avenue • Bohemia, New York 11716-1796 • USA 

Toll Free: 1-800-878-

9070 (US only) • Tel: +1 631 589 3800 • Fax: +1 631 589 3815 

http://www.koehlerinstrument.com • e-mail: [email protected] 

Petroleum Testing & Analysis Instrumentation • Custom Design & Manufacturing 

 

 

 
 
 
 
 
 
 
 
 

 

 

 

 
 
 

K60094  

Portable Heated Oil Test Centrifuge

Summary of Contents for K60094

Page 1: ...enue Bohemia New York 11716 1796 USA Toll Free 1 800 878 9070 US only Tel 1 631 589 3800 Fax 1 631 589 3815 http www koehlerinstrument com e mail info koehlerinstrument com Petroleum Testing Analysis...

Page 2: ......

Page 3: ...th the applicable standards set forth in this certification Specifications ASTM D96 API MPMS Chapter 10 4 API 2542 This unit is tested before it leaves the factory to ensure total functionality and co...

Page 4: ......

Page 5: ...ing 3 3 3 Installation 4 3 4 Accessories for Running Tests 4 4 Descriptions 4 4 1 Portable Oil Test Centrifuge 4 4 2 Control Panel 5 4 3 Temperature Controller 5 5 Operation 6 5 1 Power Connection 6 5...

Page 6: ...ler the primary focus of our business is providing you with the full support of your laboratory testing needs Our products are backed by our staff of technically knowledgeable trained specialists who...

Page 7: ...teration of this equipment from that of factory specifications is not recommended and voids the manufacturer warranty product safety performance specifications and or certifications whether specified...

Page 8: ...be 6 marked in 200 parts K60002 03058 PTFE cushion for Short Tubes K61111 Cork Stopper for Long Short and Pear Tube 4 Descriptions 4 1 Portable Oil Test Centrifuge Figure 3 Portable Oil Test Centrifug...

Page 9: ...he main chamber or bowl of the centrifuge Refer to section 4 3 for detailed description of the Temperature Controller 5 Pre heater Power Controls power to the two preheating chambers These chambers ar...

Page 10: ...re 2 With the aid of the two left most non heated centrifuge tube holders fill each of two centrifuge tubes to exactly the 50 mL 100 part mark with the test sample 3 Fill each tube with solvent to exa...

Page 11: ...ntrifuge comes to rest verify the temperature Do not disturb the oil water interface with the thermometer The test is invalid if the final temperature after centrifugation is below 52 C 125 F NOTE If...

Page 12: ...sample tube cushion inside the carousel shields Replace if worn excessively ALWAYS check moving parts both mechanical and electrical for wear and stress Replace if necessary 6 2 Replacement Parts Som...

Page 13: ...K60094 Portable Heated Oil Test Centrifuge Operation and Technical Manual K60094 Manual 9 7 Wiring Diagram...

Page 14: ...or at its sole discretion replace the equipment without charge F O B its plant provided the equipment has been properly installed operated and maintained Koehler Instrument Company must be advised in...

Page 15: ...K60094 Portable Heated Oil Test Centrifuge Operation and Technical Manual K60094 Manual 11 Notes...

Page 16: ...K60094 Portable Heated Oil Test Centrifuge Operation and Technical Manual K60094 Manual 12 Notes...