Koehler K463XX Operation And Instruction Manual Download Page 11

K463XX Cloud and Pour Point Bath 

Operation and Instruction Manual 

 

K463XX-Manual 

- 9 - 

1.3  Instrument Specifications 

 

Models: 

K46300 
K46395 
K46396 

Electrical 

Requirements: 

 
115V 60Hz 17.7 A 
220-240V 50Hz 
9.7A 
220-240V 60Hz 
9.7A 

Temperature Range: 

Ambient to -51°C 
(-60°F) 

Temperature Control 

Stability: 

 
Exceeds ASTM 
Standards 

 

Capacity: 

4 position in each 
chamber 

Bath Medium: 

Denatured ethanol, 
methanol or other 
suitable medium 

Bath Medium 

Capacity: 

 

3.79L (1.0 gal) for 
each chamber 

Altitude: 

Rated for use 
below 2000m 

Environmental 

Conditions:  

 
As per section 
1.4.1 of IEC 61010 

Laboratory Operating 

Temperature:  

 
21.11 to 22.78°C 
(70 to 73°F) 
 
 

 
 
 
 
 
 
 
 
 
 
 
 
 

Safety Information and Warnings 

 

Safety  Considerations. 

The  use  of  this 

equipment may involve 

hazardous 

materials and 

operations.  This  manual  does  not  purport  to 
address  all  of  the  safety  problems  associated 
with  the  use  of  this  equipment.  It  is  the 
responsibility  of  any  user  of  this  equipment  to 
investigate,  research,  and  establish  appropriate 
safety  and  health  practices  and  determine  the 
applicability of regulatory limitations prior to use. 

 
Equipment  Modifications  and  Replacement 
Parts. 

Any  modification  or  alteration  of  this 

equipment  from  that  of  factory  specifications  is 
not  recommended  voids  the  manufacturer 
warranty, 

product 

safety, 

performance 

specifications,  and/or  certifications  whether 
specified  or  implied,  and  may  result  in  personal 
injury  and/or  property  loss.  Replacement  parts 
must be O.E.M. exact replacement equipment. 
 

Unit  Design. 

This  equipment  is  specifically 

designed  for  use  in  accordance  with  the 
applicable  standard  test  methods  listed  in 
section  1.2  of  this  manual.  The  use  of  this 
equipment  in  accordance  with  any  other  test 
procedures,  or  for  any  other  purpose,  is  not 
recommended 

and 

may 

be 

extremely 

hazardous. 
 

This  unit  is  NOT  explosion-proof

  and  proper 

precautionary  measures  should  be  taken  during 
operation.  Avoid 

using 

this 

unit 

in  an 

environment  with  flammable  or  explosive 
vapors. 
 

Chemical  Reagents  Information. 

Chemicals 

and  reagents  used  in  performing  the  test  may 
exhibit  potential  hazards.  Any  user  must  be 
familiarized  with  the  possible  dangers  before 
use.  We  also  recommend  consulting  the 
Material Data and Safety Sheet (MSDS) on each 
chemical  reagent  for  additional  information. 
MSDS  information  can  be  easily  located  on  the 
internet 

at 

http://siri.uvm.edu 

or 

http://www.sigma-aldrich.com

. 

 
 
 
 
 
 
 

Summary of Contents for K463XX

Page 1: ...nue Bohemia New York 11716 1796 USA Toll Free 1 800 878 9070 US only Tel 1 631 589 3800 Fax 1 631 589 3815 http www koehlerinstrument com e mail info koehlerinstrument com Petroleum Testing Analysis I...

Page 2: ......

Page 3: ...K463XX Cloud and Pour Point Bath Floor Model Operation and Instruction Manual K463XX Manual 1...

Page 4: ...K463XX Cloud and Pour Point Bath Floor Model Operation and Instruction Manual K463XX Manual 2...

Page 5: ...ance with the applicable standards set forth in this certification Specifications ASTM D97 ASTM D2500 ASTM D5853 ASTM D6074 ASTM D6158 IP 15 IP 219 ISO 3015 ISO 3016 DIN 51597 FTM 791 201 NF T 60 105...

Page 6: ...K463XX Cloud and Pour Point Bath Floor Model Operation and Instruction Manual K463XX Manual 4...

Page 7: ...information on subsequent processing or recycling prior to disposal of the equipment http www dpa system dk en WEEE Products Private or professional use Responsibility for Registration and Annual Rep...

Page 8: ...ately collected WEEE Mercury containing components such as switches or backlighting lamps compact fluorescent lamps CFL Batteries Printed circuit boards if the surface of the printed circuit board is...

Page 9: ...IFICATIONS 9 2 SAFETY INFORMATION AND WARNINGS 9 3 GETTING STARTED 10 3 1 PACKING LIST 10 3 2 UNPACKING 10 4 DESCRIPTIONS 11 4 1 INSTRUMENT CONTROLS 11 4 2 ADDITIONAL ACCESSORIES 11 4 3 TEMPERATURE CO...

Page 10: ...our staff of technically knowledgeable trained specialists who are experienced in both petroleum products testing and instrument service to better understand your requirements and provide you with the...

Page 11: ...pment Modifications and Replacement Parts Any modification or alteration of this equipment from that of factory specifications is not recommended voids the manufacturer warranty product safety perform...

Page 12: ...floor Individuals must lift in accordance to proper technique See Figure below Unwrap the unit and wipe the shell with a soft cloth to remove smudges caused by wrapping material Very high ambient tem...

Page 13: ...probe 2 Set Point Temperature Display The lower green LED display shows the set point temperature of the controller 3 Programming Key Permits scrolling through controller menu parameters One Level Fo...

Page 14: ...he heating parameters needed for any particular set point temperature This operation should be done when installing a new unit after replacing or changing the bath medium type or utilizing a different...

Page 15: ...ved because there is a possibility of electrical shock 4 Proceed to test in accordance with published test methods see Section 1 3 6 Maintenance WARNING Disconnect power to the unit before servicing t...

Page 16: ...K463XX Cloud and Pour Point Bath Operation and Instruction Manual K463XX Manual 14 7 Wiring Diagram...

Page 17: ...ument Company must be advised in writing of the malfunction and authorize the return of the product to the factory The sole responsibility of Koehler Instrument Company and the purchaser s exclusive r...

Page 18: ...K463XX Cloud and Pour Point Bath Operation and Instruction Manual K463XX Manual 16 Notes...

Page 19: ...K463XX Cloud and Pour Point Bath Operation and Instruction Manual K463XX Manual 17 Notes...

Reviews: