Koehler K13902 Operation And Instruction Manual Download Page 16

K13992 Cleveland Open Cup Flash Point Tester 

Operation and Instruction Manual 

 

 

 

K13992-Manual

    

 

-

12-

 

 

6  Safety Features 

 

The Koehler K13992 Cleveland Open Cup Flash 

Point  Tester  is  equipped  with  several  safety  and 
protection  features,  which  are  described  in  the 
following sections. 
 

6.1  Over-Power Protection 

 

The Koehler K13992 Cleveland Open Cup Flash 

Point  Tester  is  equipped  with  Over-Power 
Protection  circuitry,  which  prevents  the  unit  from 
unsafe electrical conditions. If power to the unit  is 
lost, then turn off the main power and turn it back 
on again. The main power switch also functions as 
a circuit breaker. 

 

 

WARNING:

 

Disconnect  power  to  the  unit 

before  servicing  and  accessing  any  internal 
portion of the instrument to avoid exposure to high 
voltages  and/or  temperatures  which  may  result  in 
personal injury or death. If you have any questions 
about maintaining your equipment, then please do 
not  hesitate  to  contact  the  Koehler  technical 
service department. 

 

7  Maintenance 

 

WARNING.

 

Disconnect  power  to  the  unit 

before  servicing  to  avoid  exposure  to  high 
voltages  and/or  temperatures  which  may  result  in 
personal injury or death. If you have any questions 
about  maintaining  your  equipment,  please  do  not 
hesitate  to  contact  the  Koehler  technical  service 
department. 
 

7.1  Routine Maintenance 

 

The  K13992  Cleveland  Open  Cup  Flash  Point 

Tester  requires  little  routine  maintenance  to 
provide  many  years  of  continuous  service. 
However,  over  the  course  of  time,  some 
instrument  parts  may  need  to  be  replaced.  When 
ordering  replacement  part(s),  please  provide  the 
model  number,  serial  number,  and  product 
shipment  date  of  your  equipment  so  that  we  can 
ensure  you  will  receive  the  proper  replacement 
part(s). 
 

7.2  Instrument Cleaning 

 

 

To  clean  the  instrument’s  exterior,  which 
includes all painted surfaces, either a solution 
of  soap  and  water  or  laboratory  grade 
detergent may be used.  

 

  Apply cleaner to clean wipe or cloth, not to the 

instrument directly. Wipe surface clean. 

 

 

Do  Not

  clean  bath  exterior  with  organic 

chemicals such as Acetone, Toluene, Hexane, 
etc.  

 

  For more difficult cleaning of finished surfaces, 

a  dilute  solution  or  isopropanol  in  water  may 
be used. 

 

  It  is  not  recommended  that  more  aggressive 

solvents  be  used  on  painted  surfaces  since 
paint color  will  tarnish  or  be  stripped  from  the 
instrument.  

 

  Stainless  Steel  surfaces,  such  as  on  the  top 

plate,  may  be  cleaned  using  a  more 
aggressive  solvent  such  as  a  stainless  steel 
cleaner.  

 

SHOCK  AND  BURN  HAZARD:

  Only  clean 

inside  the  bath  when  equipment  is  de-energized 
and  unplugged  from  the  mains  power  supply. 
Allow adequate time for heating coils to completely 
cool before cleaning.  
 

7.3  Replacement Parts 

 

Part Number 

Description 

225-230-002 

Brick Heater, 1000W, 230V  
 

K14000 

Brass Test Cup 

 

K13900-03120  Flame Application Assembly 

 

036-108-01B 

Needle Valve, Brass, 1/8 FPT 

K13900-03117  Gas Inlet Tubing 

271-010-014 

Circuit Breaker, 2 Pole, 10A 

K138-0-9 

Burner Handle 
 

K13900-03115  Thermometer Holder  

 

K13900-03104  Heating Plate 

 

K138-1-17 

Insulation Plate 
 

010-230-004 

Wattstat, 230V 
 

K160-1-2A 

Flame Guide 

K162-0-2 

Heater Clips 

Summary of Contents for K13902

Page 1: ... Bohemia New York 11716 1796 USA Toll Free 1 800 878 9070 US only Tel 1 631 589 3800 Fax 1 631 589 3815 http www koehlerinstrument com e mail info koehlerinstrument com Petroleum Testing Analysis Instrumentation Custom Design Manufacturing K13902 K13992 Cleveland Open Cup Flash Point Tester ...

Page 2: ......

Page 3: ...pplicable standards set forth in this certification Specifications ASTM D92 ASTM D6074 AASHTO T48 ASN Z 11 6 IP 36 ISO 2592 DIN 51376 FTM 791 1103 FTM 141 4294 This unit is tested before it leaves the factory to ensure total functionality and compliance to the above specifications and ASTM standards Test and inspection records are on file for verification Jesse Kelly Application Engineer Koehler I...

Page 4: ......

Page 5: ...K13992 Cleveland Open Cup Flash Point Tester Operation and Instruction Manual K13992 Manual 1 ...

Page 6: ...K13992 Cleveland Open Cup Flash Point Tester Operation and Instruction Manual K13992 Manual 2 ...

Page 7: ...c information on subsequent processing or recycling prior to disposal of the equipment http www dpa system dk en WEEE Products Private or professional use Responsibility for Registration and Annual Reporting Koehler will not sell directly to end users in the EU and so has no responsibility to register within each EU state and to make annual reports Koehler declares that this responsibility is born...

Page 8: ...rately collected WEEE Mercury containing components such as switches or backlighting lamps compact fluorescent lamps CFL Batteries Printed circuit boards if the surface of the printed circuit board is greater than 10 square centimeters about 4 sq in Toner cartridges liquid and pasty as well as color toner Chlorofluorocarbons CFC hydrochlorofluorocarbons HCFC or hydrofluorocarbons HFC hydrocarbons ...

Page 9: ... 2 Additional Accessories Required for Testing 8 3 3 Unpacking 8 4 Setup 9 4 1 Instrument Descriptions 9 5 Operation 11 5 1 Test Procedure 11 6 Safety Features 12 6 1 Over Power Protection 12 7 Maintenance 12 7 1 Routine Maintenance 12 7 2 Instrument Cleaning 12 7 3 Replacement Parts 12 8 Wiring Diagrams 13 8 1 220 240V Unit Wiring 13 9 Troubleshooting 14 9 1 Unit does not power up 14 9 2 Unit is ...

Page 10: ... You can depend on Koehler for a full range of accurate and reliable instrumentation as well as support for your laboratory testing programs Please do not hesitate to contact us at any time with your inquiries about equipment tests or technical support Toll Free 1 800 878 9070 US only Tel 1 631 589 3800 Fax 1 631 589 3815 Email info koehlerinstrument com http www koehlerinstrument com 1 2 Recommen...

Page 11: ...ed with the use of this equipment It is the responsibility of any user of this equipment to investigate research and establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use Equipment Modifications and Replacement Parts Any modification or alteration of this equipment from that of factory specifications is NOT recommended because it v...

Page 12: ...g for correct orientation of instrument e g This Side Up 3 Carefully open top of box with box cutter and remove packing foam insert 4 Extract instrument and place on suitable cart for transportation to work area lab bench WARNING Be sure two or more individuals are available for extracting and lifting instrument from box to cart and from cart to bench Individuals must lift in accordance to proper ...

Page 13: ... or blow out the test flame as this will affect results of the test Gas Supply Connect the gas inlet to a regulated low pressure gas supply 0 5 to 1 0 psi Use only propane LPG or natural gas Do not use direct unregulated pressure from an LPG tank NOTE Be sure Flame Control Switch Figure 1 Item 3 is in the OFF position when Test Flame Applicator is not in use Power Connect the line cords to properl...

Page 14: ...per test method specifications 6 Thermometer Holder This holder is used to place the thermometer in the test cup to measure the temperature of the test specimen The holder can be tilted back to a position away from the test cup when instrument is not performing a test 7 Cleveland Open Flash Cup The brass test cup contains the test specimen and is built with a heat resistant handle An indicator lin...

Page 15: ...ases 5 to 17 C 9 to 30 F min 10 When the test specimen temperature is approximately 56 C 100 F below the expected flash point decrease the rate so that the temperature rise during the last 28 C 50 F before the flash point is 5 to 6 C 9 to 11 F min 11 Apply the test flame when the temperature of the test specimen is approximately 28 C below the expected flash point and each time thereafter at a tem...

Page 16: ... over the course of time some instrument parts may need to be replaced When ordering replacement part s please provide the model number serial number and product shipment date of your equipment so that we can ensure you will receive the proper replacement part s 7 2 Instrument Cleaning To clean the instrument s exterior which includes all painted surfaces either a solution of soap and water or lab...

Page 17: ...K13992 Cleveland Open Cup Flash Point Tester Operation and Instruction Manual K13992 Manual 13 8 Wiring Diagrams 8 1 220 240V Unit Wiring ...

Page 18: ...e consistent with an indoor laboratory environment This testing equipment should not be subjected to extremes of temperature and or moisture This equipment was shipped from the factory in a corrugated cardboard container If long term storage is anticipated re packing the instrument in a water resistant container is recommended to ensure equipment safety and longevity 12 Warranty We at Koehler woul...

Page 19: ... when you return the material to our plant Shipping containers without an RA number prominently displayed will be returned to the sender Goods must be returned freight prepaid Returns will be subject to a restocking charge the application of which will depend upon the circumstances necessitating the return Some returns cannot be authorized including certain products purchased from outside vendors ...

Page 20: ...K13992 Cleveland Open Cup Flash Point Tester Operation and Instruction Manual K13992 Manual 16 Notes ...

Page 21: ...K13992 Cleveland Open Cup Flash Point Tester Operation and Instruction Manual K13992 Manual 17 Notes ...

Page 22: ...K13992 Cleveland Open Cup Flash Point Tester Operation and Instruction Manual K13992 Manual 18 Notes ...

Reviews: