10. Press and release the Replenishment Calibration Switch on the side wall of the
Processor Tank.
11. Measure and record the volume of replenishment delivered by the system.
12. Dispose of the solution in the Graduated Cylinder.
13. Do Steps 8 through 11 at least 2 more times.
14. Determine the average volume of replenishment delivered.
15. Install the replenishment hose into the Fixer or Developer Filter Mandrel.
16. Carefully push the Processor back into the Multiloader.
Warning
See Step 9 on 4 –11.
17. Press UPDATE CAL VOL.
18. Press
∆
or
∇
key until the volume of replenishment displayed equals the average
volume determined in Step 14 on Page 5 –73.
H186_1087AC
Switch
Replenishment
Calibration
H186_1087ACB
H186_8587AC
RETURN
CAL VOL
DONE/
UPDATE