KODAK RVG 6500 System_User & Installation Guide (SM728)_Ed01
61
Information Messages
An information
“I”
error code with a message is displayed on the popup on your computer screen.
The following table lists the information messages and the action to take.
Table 3
Information Messages
Information
Error Code
Information Message
Action
I 1
Communication problem with the wireless RVG
sensor. Turn off then on your sensor or restart
your computer. If the problem persists, contact
your representative.
〈
Turn Off then ON the RVG sensor.
〈
Or reboot your computer.
〈
If the problem persists, contact your
representative.
I 2
Wireless sensor error. Turn off then on your
sensor and restart the acquisition software. If
the problem persists, contact your
representative.
1.
Turn Off then ON the RVG sensor.
2.
Click
to access the
RVG
Acquisition
interface
.
If the problem persists, contact your
representative.
I 3
Error during image acquisition. Turn off then on
your wireless sensor and restart the acquisition
software. If the problem persists, contact your
representative.
1.
Turn Off then ON the RVG sensor.
2.
Click
to access the
RVG
Acquisition
interface
.
If the problem persists, contact your
representative.
I 4
Memory allocation problem. Turn off then on your
wireless sensor. If the problem persists, contact
your representative.
Turn Off then ON the RVG sensor.
If the problem persists, contact your
representative.
I 5
Wireless sensor has storage/transmission
memory problem. Contact your representative.
Contact your representative.
I 6
Incorrect IPS calibration. Contact your
representative
Contact your representative.
I 7
IPS Batteries have run down. Change the
batteries.
Change the batteries of the IPS aiming
ring.
I 9
Writing transmission error. Turn off then on your
wireless sensor and restart the process. If the
problem persists, contact your representative.
1.
Turn Off then ON the RVG sensor.
2.
Redo the process
If the problem persists, contact your
representative.
I 10
Reading transmission error. Turn off then on your
wireless sensor and restart the process. If the
problem persists, contact your representative.
1.
Turn Off then ON the RVG sensor.
2.
Redo the process
If the problem persists, contact your
representative.
Summary of Contents for RVG 6500
Page 1: ...KODAK RVG 6500 System and KODAK RVG 6500 IPS System User Installation Guide...
Page 6: ...2 Chapter 1 Conventions in This Guide...
Page 8: ...4 Chapter 2 KODAK RVG 6500 Systems Packaging...
Page 12: ...8 Chapter 3 KODAK RVG 6500 Systems Overview...
Page 20: ...16 Chapter 4 Imaging Software Overview...
Page 34: ...30 Chapter 5 Setting Up the KODAK RVG 6500 Systems...
Page 42: ...38 Chapter 6 Acquiring an Image...