DIAGNOSTICS
Error Codes
10DEC05
DG4825-1
Page
95 of 180
Calibration
16001 - Calibration Error: Did not find START OF SCAN position
16002 - Calibration Error: Did not find Offset
16003 - Calibration Error: Did not find Amplitude
Description: During the calibration, the CR 825/850 SYSTEM did not locate the
“START OF SCAN” position within the given time.
Cause 1: Calibration of the CR 825/850 SYSTEM is not correct.
Action:
Calibration for the CR 825/850 SYSTEM
Cause 2: The alignment of the optics is not correct.
Action:
Do:
•
adjustment for the
•
adjustment of the
Cause 3: The SLOW SCAN CONTROLLER BOARD A6 malfunctioned.
Action:
Do the checkout for the
SLOW SCAN CONTROLLER BOARD A6 and COIL
Description: The GALVO did not detect the “Offset”.
Cause 1: The exposure of the PLATE is not correct.
Action:
1.
Use the DOSIMETER to check the exposure of the PLATE.
2.
Check for scratches on the edges of the SCREEN.
Cause 2: The GALVO is not aligned correctly.
Action:
Do:
•
adjustment of the
•
checkout for the
Description: The GALVO BOARD A4 did not detect the “Amplitude”.