DIAGNOSTICS
Error Codes
10DEC05
DG4825-1
Page
100 of 180
16021 - Malfunctioning PMT1: Gain out of range
16022 - Malfunctioning PMT2: Gain out of range
Description: For a given high voltage, the low gain of PMT1 might prevent it from
reaching the correct signal level.
Cause 1: The exposure on the PLATE is not correct during the calibration.
Action:
Calibration for the CR 825/850 SYSTEM
Cause 2: The alignment of the optics is not correct.
Action:
Do:
•
adjustment for the
•
adjustment of the
Cause 3: The PMT malfunctioned.
Action:
Install a new PMT. Use
.
Cause 4: The PMT/DAS BOARD A5 malfunctioned.
Action:
Do the checkout for the
Description: For a given high voltage, the low gain of PMT2 might prevent it from
reaching the correct signal level.
Cause 1: The exposure on the PLATE is not correct during the calibration.
Action:
Calibration for the CR 825/850 SYSTEM
Cause 2: The alignment of the optics is not correct.
Action:
Do:
•
adjustment for the
•
adjustment of the
Cause 3: The PMT malfunctioned.
Action:
Install a new PMT. Use
.