SM731_K2200_01_en
06/2009
5-2
Cleaning
Clean the outside of the system with a damp paper towel or soft cloth using
an alcohol-based, non-corrosive cleaner.
Disinfecting
If necessary, wipe off surfaces with disinfectant.
CAUTION:
•
Do not allow liquids to drip into the system.
•
Do not spray cleaner or disinfectant directly onto the machine.
•
Protect the system from contamination using barriers available
from dental distributors.
•
Follow the manufacturer’s safety recommendations when using
the cleaner or disinfectant.
Error Messages
Error Message
Cause
How to cancel
I 01 and a red LED
lights up on the
control unit
Cooling cycle: this message can appear
during a period of intensive use.
Do not turn off the system
. The error
message will disappear when the
system returns to a satisfactory
temperature.
CAUTION:
If you turn off power to the system, the microprocessor does not calculate the cooling time, and for
safety reasons considers that the system has not gone into the cooling cycle.
I 02
Request for a x-ray tube seasoning
Refer to Installation Manual
E 01
plus audible alarm
Release of the radiography control button
before the end of the exposure. The display
shows the remaining exposure time (Based
on this time, decide whether to develop the
film or make another exposure).
Press the selector knob to stop the
alarm.
E 02
The radiography control was activated while
the unit was being powered on.
E 03-E 04
Problems with the exposure time control.
E 10 to E 18
kV voltage error.
E 20 to E 24
Filament voltage error.
E30
Problem with voltage to main power supply
or to chemical capacitor.
E 40 to E 46
System error (problems with the
microprocessor on the power board).
E 50 to E 54
Problems with the I2C bus (the connection
between the control panel and the power
board).
Turn off the system and restart. If the
problem persists, call a qualified service
technician and discontinue using the
equipment.
Table 3: Error Messages