MAS
page 12
MAS K05/0722
10. Totalizer
The totalizer is designed to provide a totalizer function for the MAS-11/MAS-31. It
will display the totals as well as the flow.
Display:
The totalizer display has three screens which are accessed by pressing the
button as indicated below:
Screen 1 (start-up screen): Flow units are shown together with the actual flow
Screen 2:
Totalizer
Screen 3:
Actual flow is shown together with the totalizer
Totalizer:
A total of 8 digits can be shown by the totalizer ranging from .0000001 to
99999999. The decimal point will automatically shift position as the total
increases. Upon reaching the maximum count (99999999), the totalizer will “roll-
over” be cleared and counting starts from zero again.
The totalizer is cleared by going to screen 2 or 3 and then press the button for
more then 5 seconds. The clearing of the totalizer can be observed on the
screen.
The total count is stored in non-volatile memory every 5 minutes. If the unit is
switched off within 5 minutes from power-up then no total will be saved and the
previous total will be shown at the next power-up.
Due to the combination of the hardware/LCD it may happen that
during power up the screen remains blank. Please turn the unit off
and then back on again.