8. Operation / Configuration / Adjustments
8.1 Leak testing
Before starting up, all pressure lines, fittings and other components of the
pressurized system circuit must be tested for leaks.
Test methods:
- pressure drop method
- shut off the line circuit, pressure drop observed on time
and / or
- method of blowing
- spraying relevant surfaces by means soap bubbles
medium - a possible blowing magnification is observed
8.2 Permissible vibration load at the mounting point
The instruments should always be installed in locations free from vibration.
If necessary, it is possible to isolate the instrument from the mounting point
by installing a flexible connection line between the measuring point and the
pressure gauge and mounting the instrument on a suitable bracket.
If this is not possible, the following limit values must not be exceeded:
Dry gauges:
Frequency range < 150 Hz Acceleration < 0.7 g (7 m/s2)
Liquid-filled gauges: Frequency range < 150 Hz
Acceleration < 4 g (40 m/s2)
The liquid filling must be checked on a regular basis.
The liquid level must not drop below 75 % of the gauge diameter.
Summary of Contents for MAN-D Series
Page 1: ...Operating Instructions for Diaphragm Differential Pressure Gauges Model MAN D...
Page 9: ...MAN D MAN D K02 0522 page 9 11 Technical Information...
Page 10: ...MAN D page 10 MAN D K02 0522...
Page 11: ...MAN D MAN D K02 0522 page 11...
Page 15: ...MAN D MAN D K02 0522 page 15 Option Bracket for wall or pipe mounting...