5. Unit Check-up
These devices are checked before shi
pment and sent away in perfect condition. Should
the damage to a device be visible, we recommend a thorough inspection of the delivery
packing. In case of damage, please inform your parcel service/ forwarding agent
immediately, since they are responsible for damages occuring during transit.
Scope of delivery
Standard scope of delivery applies:
Flow meter
Connection cable with plug
: KF-7000
Instruction manual
6. Mechanical Connection
Before
Installation
Please ensure that the actual flow volume is in agreement with the measuring
range of the device. The reference number of measuring range can be read
from the item number label.
Note: Continuously exceeding the measuring range can cause damage
to bearings.
Make sure that the maximum operational pressure and temperature are not
exceeded.
Ensure that there is no remaining packing material inside the unit.
KFG:
The calibration of the unit is performed in horizontal position; the item label faces
upwards. With other mounting positions, a measured-value deviation of 0.25 %/°C at
50% nominal flow should be taken into consideration.
KFF/KFG:
Mount the measuring device on desired position. Two mounting flanges are
supplied for this purpose.
Connect unit to your piping system.
Do not remove the meter fittings as the calibration may change.
Attention:When tightening fittings, it is necessary to work with two suitable
wrenches. A transfer of torque on device's housing while tightening
the fittings can result in the destruction of the measuring device.
Check the sealing of
all connections.
KFF, KFG
Page 4
03/04