HND-T205 

HND-T205 K05/0422 

 

page 13 

7.8  Operation of Logger 

The device supports two different logger functions: 

 

„Func-Stor“:

 

each time when „store“ (key 6) is pressed a measurement will 
be recorded. 

„Func-CYCL“:

  measurements will automatically be recorded at each interval, 

which  was  set  in  the  logger  menu  ‚CYCL‘  until  the  logger  will 
be  stopped  or  the  logger  memory  is  full.  The  recording  is 
started by pressing „Store“ 2 seconds. 

The logger records 1 measurement result each time  
The data set consists of: 

-measuring value at time of recording 

   

 

 

 

-time and date of recording 

 
For the read out and evaluation of the data the software HND-Z034 (at least V. 
1.7) has to be used. The software also allows easy configuration and starting of 
the logger. 
When  the  logger  is  activated  (Func  Stor  or  Func  CYCL)  the  hold  function  is  no 
longer available, key 6 is solely used for the operation of the logger functions. 
 

7.8.1 

Storing Single Measurements („Func-Stor“) 

a)  Record Measuring: 

If  the  logger  function  “Func  Stor”  was  activated  (see  chapter  Device 
configuration), up to 99 measuring data sets can be recorded manually.  
 
 

Press shortly: data set will be recorded (“St.XX” will be shown shortly. 
XX is the number of the data set) 

If the logger memory is full, the display will show:  
 

b)  Viewing Recorded Measurings: 

Stored data sets can be either read out via PC Software HND-Z034, or be viewed 
in the display directly. 

 

Press for 2 seconds:

 The display will show:  

 
 

Note: “rEAd Logg” only appears, if there are already data stored! If 

memory is empty, the configuration menu 

will show. 

 

 

Summary of Contents for HND-T205

Page 1: ...Operating Instructions for Precision Hand Held Thermometer Model HND T205...

Page 2: ...6 Special Functions 9 7 7 Probe connection 12 7 8 Operation of Logger 13 7 9 Some Basics of Precision Temperature Measuring 15 7 10 Fault and System Messages 17 7 11 Calibration Services 17 8 Maintena...

Page 3: ...nly to be used maintained and serviced by persons familiar with these operating instructions and in accordance with local regulations applying to Health Safety and prevention of accidents When used in...

Page 4: ...nd zero point increase entry for example 6 Electrical Connection 6 1 Mains Operation Attention When using a power supply unit please note that operating voltage has to be 10 5 to 12 VDC Do not apply o...

Page 5: ...her devices Internal connection in third party devices e g connection GND and earth may result in not permissible voltages impairing or destroying the device or another device connected Warning Operat...

Page 6: ...olated interface adapter accessory HND Z031 Operation as analogue output Connection via suitable cable Attention The output mode has to be configured p r t 7 6 6 and influences battery life 2 Probe co...

Page 7: ...he min Value will be deleted Key 6 Measurement Hold current measuring value HLD in display or operation of logger functions p r t chapter 7 8 7 5 Device Configuration Note Some menu items will be show...

Page 8: ...e will be automatically switched off as soon as this time has elapsed if no key is pressed no interface communication takes place oFF Power off function inactive continuous operation e g mains operati...

Page 9: ...besides the standard conversion of resistance to temperature following EN60751 Lin E 751 also other curves can be used The user sensor curve can be read and edited by configuration software The standa...

Page 10: ...ard curve LIN E 751 and the user sensor curve Lin USEr posses separate correction settings 7 6 5 Average filter tAuG The filter forms the arithmetic floating average over the entered time The displaye...

Page 11: ...on 23 read max alarm rail AL AL Hi 216 read zero displacement 32 read configuration flag alarm function 1 Alarm horn 3 BitLoggerOn 50 BitCyclicLogger 51 217 set zero displacement 222 Read power off ti...

Page 12: ...urned interface message 7 6 8 Real Time Clock CLOC The real time clock is used for the logger function Recorded values are also containing the point of time when they were measured Please check the se...

Page 13: ...tware also allows easy configuration and starting of the logger When the logger is activated Func Stor or Func CYCL the hold function is no longer available key 6 is solely used for the operation of t...

Page 14: ...hapter 8 Device configuration Max number of measurings 16384 a Starting a recording press for 2 seconds the recording will be started Each recording is signalled with a short display of St XXXXX XX is...

Page 15: ...03 0 0005 temperature A 0 15 0 002 temperature Error over measuring range Error over range 50 150 C For applications demanding higher precision than given by these classes we suggest to adjust the dev...

Page 16: ...probe or the probe can have a better heat flow to the ambient temperature as to the objects surface Therefore specially designed surface probes should be used The measuring precision depends mainly o...

Page 17: ...to manufacturer for repair Sensor error no sensor connected Connect sensor to socket sensor cable or device defective return to manufacturer for repair Err 1 Value exceeding measuring range Check Is t...

Page 18: ...99 99 199 99 C or 200 0 850 0 C Fahrenheit values accordingly Accuracy at nominal temperature 25 C 0 03 C in the range 199 99 199 99 C 0 1 C 1 digit in the range 200 0 850 0 C Resolution 0 01 C or 0 1...

Page 19: ...0 min can be deactivated Zero point and increase entry zero point and increase correction can be entered digitally Logger functions Manual 99 datasets Cyclic 16 384 datasets Adjustable cycle time 1s 1...

Page 20: ...50 400 C approx 10 sec HND TF02 Immersion probe for liquids and gases 4 wire Like HND TF01 but with 1 10DIN Class B 0 03 C at 0 C and flexible sheath tube 3 mm 50 400 C approx 10 sec HND TF03 Insertio...

Page 21: ...gg crate foam 275x229x83 mm HND Z022 Large case with recess 394x294x106 mm HND Z023 Interface converter on RS232 galvanically isolated HND Z031 Interface converter on USB galvanically isolated HND Z03...

Page 22: ...elow EN 61326 1 2013 Electrical equipment for measurement control and laboratory use EMC requirements Part 1 General requirements EN IEC 63000 2018 Technical documentation for the assessment of electr...

Reviews: