background image

 
 
 
Heinrichs Messtechnik GmbH 

EX-Supplementary Operating Manual  
HPC Mass Flow Meter 

Page 13 of 20 

 

5

 

Installation and commissioning 

5.1

 

Mounting 

The mounting instructions for the standard sensor also apply to the explosion-proof sensor. This handbook 
complements that of the standard sensor. 
The sensor may only be separately installed and connected with a special sensor cable, whereby the 
maximum CL and LL of the connection cable must be observed. 

 
Refer to the standard operating manual for the connection scheme of the sensor 

5.2

 

General conditions for safe operation 

a)

 

If the conditions described in this document are not adhered to or if there is any inappropriate 

interference with the equipment, all the manufactures warranties expire. 

b)

 

Conditions described in this manual, as well as the permitted operating conditions which have been 

defined for the sensor and which are stated on the rating plates must be adhered to. 

c)

 

Appropriate measures shall be met to prevent any unintentional or inadmissible damage to the 

device. 

d)

 

The operator shall ensure that the equipment is only installed in areas which comply with the 

approved types of protection and environments. 

e)

 

All connected electrical equipment must be suitable for its intended use. 

f)

 

The operator shall ensure protection against lightning according to local regulations. 

g)

 

The sensor is maintenance-free. Substitution of components may impair intrinsic safety 

5.3

 

Requirements for installation in all environments 

a)

 

The installation of the intrinsically safe circuits requires a control drawing (system description), to be 

issued by the operator/erector. 

b)

 

The equipment is only to be installed and connected in a de-energized state. 

c)

 

The sensor can be installed in zone 0 as well as zone 20 or 21. The intrinsically safe circuits are to be 

installed (from specialists) according to applicable regulations. 

d)

 

Equipotential bonding between the sensor and transmitter is essential to guarantee conformity of the 

intrinsically safe circuits. 

e)

 

It is to be insured that the intrinsically safe circuits of the sensor cable are not laid together with non-

intrinsically safe circuits. 

f)

 

When using a connection cable other than the manufactures specified cable, the intrinsic safety of the 

cable is to be validated using the cables nominal parameters. 
Refer also to section 4.3.3.2 “Cable Parameters” on page 10 

g)

 

If the sensor is used at an ambient temperature below -20 °C, suitable cables are to be used. 

h)

 

The operational wall thickness of the stainless steel flow tubes are < 1 mm. Therefore, in cases where 

the sensor is operated in zone 1 atmospheres and where zone 0 may be present in tubes, the 

Summary of Contents for Heinrichs HPC Series

Page 1: ...y Operating Manual for explosion proof flow meters This operating manual contains important information for the operation in potentially explosive atmospheres Please read the instructions carefully an...

Page 2: ...nstallation 9 4 3 1 Thermally insulated sensor 9 4 3 2 Heated sensor 9 4 3 3 Connection sensor transmitter 9 4 4 EC Type Examination 11 4 4 1 Sensor Parameters 11 4 4 2 Ambient temperature range Ta 12...

Page 3: ...tion November 2018 or later These instructions are supplementary operating manual for non explosion proof Coriolis flow sensors If you do not have a copy of the latter instructions please request one...

Page 4: ...to be performed by a technician trained to work with explosion proof devices or by a Heinrichs Messtechnik service technician Warning Any safety relevant maintenance or repairs in terms of explosion p...

Page 5: ...substantial material damage or even death Always comply to these warnings and proceed with caution Caution means that failure to take the prescribed precaution could result in material damage or destr...

Page 6: ...Messtechnik GmbH Robert Perthel Strasse 9 D 50739 Cologne Germany Phone 49 221 49708 0 Fax 49 221 49708 178 Internet www heinrichs eu Email info heinrichs eu Product type Mass flow rate meter for liqu...

Page 7: ...dust mixtures are likely to occur 3G 2 3D 22 Equipment in this category is intended for use in areas in which explosive atmospheres caused by gases vapours mists or air dust mixtures are unlikely to o...

Page 8: ...sensor using a dedicated cable specially designed for this purpose Such a cable can be purchased from Heinrichs Messtechnik during your order process 4 2 Device identification The rating plates on Hei...

Page 9: ...for the fluid is indicated on the sensors rating plate It is also the operator s responsibility to ensure that no hazards are created by hot surfaces pursuant to EN 1127 1 Explosive atmospheres Explos...

Page 10: ...Electrical installations in hazardous areas for Europe Section 18 of the Canadian Electrical Code for Canada or Article NEC 504 and or NEC 505 for the United States Applicable guidelines pertaining to...

Page 11: ...35 C Da Ex ia IIIC T190 C T240 C Db and are approved for the use in potentially explosive atmospheres hazardous classified locations Warning In order to comply with the requirements for explosive atmo...

Page 12: ...e temperature and equipment protection level are dependent upon the connection type process temperature and ambient temperature as follows Table1 Temperature Classes Connection via HAN R23 connector p...

Page 13: ...le for its intended use f The operator shall ensure protection against lightning according to local regulations g The sensor is maintenance free Substitution of components may impair intrinsic safety...

Page 14: ...user installer The options are detailed under the description in the certificate When the maximum process temperature of the final installation is determined by the end user the temperature class ass...

Page 15: ...sses those aspects covered by a close inspection and in addition identifies those defects such as loose terminations which will only be apparent by opening the enclosure and or using where necessary t...

Page 16: ...in conjunction with explosive atmospheres Such equipment shall only be repaired by expert personnel trained and authorized to do so Damaged threaded or flat joints which may impair the IP protection o...

Page 17: ...osition J Enclosure options Non Ex relevant position K Type of installation Non Ex relevant position L Sensor configurations R Remote 40 C to 80 C 40 F to 176 F L Remote 40 C to 180 C 40 F to 356 F X...

Page 18: ...Heinrichs Messtechnik GmbH Page 18 of 20 EX Supplementary Operating Manual HPC Mass Flow Meter 9 Declaration of Conformity...

Page 19: ...Heinrichs Messtechnik GmbH EX Supplementary Operating Manual HPC Mass Flow Meter Page 19 of 20...

Page 20: ...Heinrichs Messtechnik GmbH Page 20 of 20 EX Supplementary Operating Manual HPC Mass Flow Meter...

Reviews: