Heinrichs Messtechnik GmbH
OPERATING MANUAL TMU-W / UMC4
Page 37 of 118
Wire designation of the sensor cable
Designation
Terminal / Pin designation
Type of protection
Ex ia
Non-Ex
(Standard)
Sensor lines
S
1
x
x
SENSOR1 -
2
x
x
S
3
x
x
SENSOR2 -
4
x
x
TIk-
5
x
x
Temperature sensor -
6
x
x
Temperature
7
x
x
TIk+
8
x
x
EXCITER1
9
x
x
EXCITER2
10
x
x
Shield
Shield
x
x
Table 8: Wire designation of the sensor cable
For the connection between the sensor and transmitter, one of the following cables is supplied:
Non-explosion proof applications SLI2Y(ST)CY 5 x 2 x 0.5 mm² grey (max. 300 m)
Explosion-proof applications
SLI2Y(ST)CY 5 x 2 x 0.5 mm²
blue (max. 300 m)
(blue for explosion-proof applications, grey for non-explosion proof applications).
Customer specific cable lengths can be ordered during the ordering process.
Attention: Cables not certified by the manufacturer may impair the accuracy of the measurement as well as
EMC compliance.
Suitable alternative cables:
SLI2Y (ST) C11YÖ 5 x 2 x 0.5 mm or
SG [5(2 LiY 0.50)St]FStC11Y