Trouble shooting
-
Device shows incorrect values: Compare process data, density, viscosity, temperature and pressure with the
values on the scale. If they deviate, convert the scale values using the VDE/VDI 3513 standard.
-
Pointer does not react in spite of varying flow: The pointer may have gotten stuck; remove the cover and
move the pointer; if the pointer can be moved easily, the float cannot move. If the pointer is unable to move
further, send the device to the head office for servicing.
-
The float is stuck at one place due to dirt: Disassemble the device. If necessary, dismantle and clean the float.
Install a magnetic filter if there are magnetic contaminants.
-
Electrical equipment are not functioning: Check the auxiliary power. Are suitable power supply equipment
connected, have the terminals been selected correctly, has the parameterization carried out correctly?