13
Safety Information
Application in hazardous locations:
Stratos A401B is approved for operation in hazardous locations.
Observe all applicable local codes and standards for the installation of
electrical equipment in hazardous locations.
For orientation, please refer to NFPA 70 (NEC), ANSI/ISA-RP12.06.01.
• When installing the device in a hazardous location, observe the
specifications of the Control Drawing (download: www.knick.de).
• Before commissioning, you must provide proof that the device may
be connected with other equipment, such as a supply unit includ-
ing cables and wires.
• In hazardous locations the device shall only be cleaned with a
damp cloth to prevent electrostatic charging.
• Devices which have already been used shall be subjected to a pro-
fessional routine test before they may be operated in another zone
or another type of protection.