104
Calibration
Calibration with Calibration Solution
Input of temperature-corrected value of calibration solution with
simultaneous display of cell factor.
Be sure to use known calibration solutions and the respective temper-
ature-corrected conductivity values (see calibration solution tables in
the appendix). During the calibration procedure the temperature must
be kept constant.
Note:
When using an ARF 210/215 flow-through fitting, you should use the
inluded calibration beakers (identical dimensions and materials) for
calibration to prevent calibration errors.
Display
Action
Remark
Select Calibration.
Press enter to proceed.
Select CAL_SOL
calibration method.
Press enter to proceed.
Ready for calibration.
Hourglass blinks.
Display (3 sec)
Now the device is in
HOLD mode.
Immerse sensor in
calibration solution.
Enter the temperature-
corrected value of the
calibration solution us-
ing the arrow keys (see
table in the appendix).
Press enter to confirm.
Lower line:
display of cell factor
and temperature