Error Messages
Error
Icon
(blinks)
Problem
Possible causes
A
lar
m c
on
tac
t
Red LED
out 1 (22 mA
)*
out 2 (22 mA
)*
ERR 03
Temperature probe
Open or short circuit
Temperature range exceeded
x
x
x
x
ERR 11
Current output 1
Current below 0 (3.8) mA
x
x
x
ERR 12
Current output 1
Current above 20.5 mA
x
x
x
ERR 13
Current output 1
Current span too small / too large
x
x
x
ERR 21
Current output 2
Current below 0 (3.8) mA
x
x
x
ERR 22
Current output 2
Current above 20.5 mA
x
x
x
ERR 23
Current output 2
Current span too small / too large
x
x
x
ERR 33
Sensocheck
Glass electrode
x
x
x
• Zero error, Sensoface active
• Slope error, Sensoface active
• Response time exceeded, Sensoface active
• Calibration interval expired, Sensoface active
* if output has been set to Alarm = 22mA during configuration