8.2 Assembly

Perform steps 1, 2 and 3 for mounting the interior body.

1.  Insert the interior body in the sensor body. Turn it 

until the slot in the interior body engages with the 

pin in the sensor body.

2.  Press the interior body in the sensor shaft until the 

stop  is  reached. Then  screw  on  the  lock  nut  and 

tighten it using a 3/8" open-end wrench.

3.  Check all O-rings for damage. Replace if necessary.

4.  Half fill the membrane module with O

2

 electrolyte. 

Make sure that there are no air bubbles in the mem-

brane module. 

Note: Air bubbles can be removed by carefully tap-

ping on the membrane module.

Caution!

The  electrolyte  is  strongly  alkaline.  Avoid 

contact  of  electrolyte  with  skin,  mucous 

membranes  or  eyes.  Wear  suitable  protec-

tive  gloves  and  goggles  when  servicing  the 

 sensor.

5.  Slip the membrane module over the interior body 

while  holding  the  sensor  in  a  vertical  position. 

 Remove excess electrolyte with a paper tissue.

Note: Make sure that there is no electrolyte, process 

solution or any other contamination between mem-

brane module and cap sleeve.

6.  Carefully  slip  the  cap  sleeve  over  the  membrane 

module and screw it tight.

7.  Note: After each replacement of electrolyte, mem-

brane  module  or  interior  body,  you  must  repolar-

ize (see section 6. Polarization) and recalibrate the 

 sensor (see section 7. Calibration).

8.1 Disassembly

1.  Unscrew the cap sleeve from the sensor shaft and 

carefully pull it off the sensor.

2.  Pull  off  the  membrane  module  from  the  interior 

body. Before replacing electrolyte, you must remove 

the membrane module from the cap sleeve.

3.  Rinse the interior body with distilled water and care-

fully dab it dry with a paper tissue.  

Perform  steps  4  and  5  only  when  dismounting  the 

 interior body:1 ⁾

4.  Loosen the lock nut using a 3/8" open-end wrench. 

5.  Pull the interior body out of the sensor shaft. (Use a 

plier if necessary.)

Note

Do not twist the interior body when pulling it out. The 

contact pins might break off.

4. Hazardous-Area Concept

The  sensors  are  intrinsically  safe  equipment  of  Cat-

egory  1. They  may  be  installed  within  explosive  gas 

atmospheres up to Zone 0.

1.  Connect the sensors only to an intrinsically safe, ap-

proved cable of Type CA/MS-***X** (BVS 09 ATEX E 

083 X) from Knick Elektronische Messgeräte GmbH 

& Co. KG.

2.  Do not operate the sensors under electrostatically 

adverse conditions. Prevent strong vapor and dust 

streams  from  acting  directly  on  the  sensors  and 

measuring cable. The shaft length is not Ex-relevant.

1.  Observe the stipulations for electrical installations in 

explosive gas atmosphere (EN/IEC 60079-14) when 

using the sensors and cables.

2.  Observe the stipulations/conditions for safe use of 

the connected intrinsically safe Memosens® sensor. 

1. Introduction

The  Knick  oxygen  sensors  of  the  SE  707  series  with 

Memosens

®

  connector  system  are  low-maintenance 

and pressure-resistant oxygen sensors with integrated 

temperature detector.

SE 707 sensors are suitable for in-line measurement 

of dissolved oxygen and temperature in the different 

fields of analytical chemistry: 

• Biotechnology

• Pharmaceutical industry

• Fermentation

SE 707 sensors are suitable for sterilization and auto-

claving and are CIP capable.

 

Instructions for Use of Series SE 707-Memosens

®

 Oxygen Sensors

Caution!

Please read these instructions carefully before 

putting  the  sensor  into  operation,  in  order 

to  ensure  faultless  function.  Operation  and 

 servicing of the sensor must be carried out by 

specially  trained  personnel  authorized  by  the 

operating company.

Warning!

After an Ex-certified sensor has been used with 

non-Ex components, this sensor must no longer 

be used in a hazardous location.

II 1 G Ex ia IIC T3/T4/T6

BVS 10 ATEX E 089 X

0044

 

 

Instructions for Use of Series SE 707-Memosens

®

 Oxygen Sensors

Knick

Elektronische Messgeräte GmbH & Co. KG

P.O. Box 37 04 15

D-14134 Berlin    

Germany

Tel:  +49 (0)30 - 801 91 - 0

Fax: +49 (0)30 - 801 91 - 200

www.knick.de

[email protected]

TA-SE707MS-KNX02 

20100709

SE 707

MEMO

SENS

Memosens

®

 connector

Pg 

13.5

Ø 12 mm

120 

/ 225 

mm

12345

    67

Serial No.

Note!

If the ambient temperatures indicated are observed, 

no  impermissible  temperatures  for  the  temperature 

class in question occur at the equipment.

Cap sleeve

Membrane

Membrane module

Interior body

Recommended fittings:

ARI 106

ARF 200/202

ARF 210

ARD 230

SensoGate® WA 130/131

2. Safety information

The sensor is intended only for the use described in 

section 1. Introduction above. Be sure to wear safety 

goggles and gloves when cleaning or calibrating a sen-

sor using acid or alkaline solutions.

The  sensors  are  designed  for  an  operating  pressure 

of  6 bars  and  meet  the  PED  requirements  (Pressure 

Equipment Directive 97/23/EC, Article 3, Sec. 3). 

3. Product description

The markings on each sensor or on the packaging label 

include the following information:

Knick 

Manufacturer of sensor

   SE 707/*-*MSN Model designation

   

Memosens® connector (MS)

   

   

Length (1: 120 mm, 2: 225 mm)

-5…80 °C  

Temperature range

The serial number is printed on and also electronically 

stored  in  the  sensor  head.  See  also  enclosed  quality 

certificate.

6. Polarization

Prior to first use or when the sensor has been discon-

nected from the voltage source for longer than 5 min-

utes, you must polarize the sensor by connecting it to 

an operating transmitter and then calibrate it.

Polarization time depends on the time without polar-

ization voltage. The respective polarization times are 

shown in the following table.

Time without
polarization voltage d [min.] 

Required
polarization time [min.]

d > 30

360

30 > d > 15

6 * d

15 > d > 5

4 * d

d < 5

2 * d

7. Calibration of sensor and transmitter 

When the sensor is continuously used, you must cali-

brate it at regular intervals. Calibration is also required 

after  replacement  of  electrolyte,  membrane  module 

or interior body.

Note: To check whether the sensor needs to be cali-

brated, you can take it in the air, dab it dry and check 

the display on the transmitter. The indicated oxygen 

saturation should be close to 100%. If not, you must 

calibrate the sensor.
7.1 One-point calibration

Calibration in air with known water-vapor saturation

1.  Take the sensor out of the process medium.   

2.  Clean the sensor shaft.

3.   Carefully dab the membrane dry with a paper tissue.

4.   Perform  the  calibration  as  instructed  in  the  trans-

mitter manual. 

(Refer to the manual for further calibration  methods.)

7.2 Two-point calibration

For  two-point  calibration,  always  start  with  the  zero 

calibration before calibrating the slope.

With  two-point  calibration,  you  achieve  a  higher  ac-

curacy with very low oxygen concentrations. For zero 

calibration, we recommend an oxygen-free medium, 

e.g. nitrogen with a purity level of at least 99.995%.

After the sensor signal has stabilized, the sensor and 

transmitter  are  calibrated  to  the  zero  value  of  the 

 desired process variable.

Please refer to the user manual of the transmitter.

8. Maintenance

The  maintenance  intervals  depend  on  the  sensor 

 application. 

Maintenance comprises cleaning the sensor and 

replacing wear parts and consumables such as:

  - O-rings

  - Electrolyte

  - Membrane module

  - Interior body

9. Cleaning the sensor

Depending on the measured medium, the sensor must 

be cleaned to ensure accurate measurements.

You can clean the sensor shaft with water and a suit-

able brush. 

Use water and a soft paper tissue to remove dirt par-

ticles from the membrane or membrane module.

After cleaning, thoroughly rinse with clean water.

Note:  Do  not  use  cleaning  agents  or  alcohol. This 

could damage the sensor or lead to fault currents.

10. Warranty 

12-month warranty after date of shipment in the event 

of manufacturing faults.

11. Specifications

Measuring range

pO2 < 1200 mbars

Detection limit

1 ppb

(3 ppb in in liquids containing CO2)

Accuracy

1 % + 1 ppb

(1% + 3 ppb in liquids containing CO2)

Response time at 25 °C 

(air –> N2)

98 % full scale < 90 s

Signal in ambient air

290 ... 500 nA

Residual signal in  

O2-free medium

≤0.1% of signal in ambient air

In CO2: ≤0.01% of signal in ambient air

Flow dependence

≤ 5%

Permissible pressure 

range (measurement)

0.2 ... 6 bars absolute

Mechanical  

pressure resistance

Max. 12 bars absolute

Permissible  

temperature range

-5 ... 80 °C

Temperature range 

(stability)

-5 ... 121 °C

O-ring material

Silicone 

(FDA & USP approved up to VI)

Membrane material

PTFE / silicone / PTFE,

FDA compliant

(steel-mesh reinforced)

Process-wetted  

sensor parts

Stainless steel, 1.4435

(material certificate 3.1B)

Surface roughness

N5 (RA < 0.4µm)

Temperature detector

NTC 22 kΩ

Interior body  

incl. sensing electrodes

Replaceable (ZU 0568)

Sensor cap

Memosens®

12. Accessories / Spare parts

O2 membrane module, separate

ZU 0563

O2 membrane kit, consisting of:

Membrane module (4x),

O-ring set (1x),

25ml electrolyte (1x)

ZU 0564

O2 electrolyte 25 ml

ZU 0565

Interior body for SE 707

ZU 0568

O-ring set, silicone, FDA

ZU 0679

1  ⁾ The  interior  body  must  only  be  replaced  if  the 

electrode system is poisoned or the interior body 

is broken.

Cap sleeve

5. Installation and commissioning

1.  On  unpacking,  check  the  sensor  for  mechanical 

damage. Report any damage to your Knick service 

team.

2.  Prior to first use, you must fill the sensor with new 

electrolyte (see section 8). 

3.  Remove the protective cap. 

4.  Please refer to the instruction manual of the relative 

fitting in order to install the sensor.

5.  Sensor and cable are plugged together and locked 

by a quarter turn of the coupling ring. See function 

description of Memosens

®

 cable.

Prior to first use, you must polarize the sensor. After 6 

hours, the sensor is polarized and ready for operation. 

See section 6. Polarization for more information.

Note

As delivered, a moderate slope value is stored in the 

sensor. The sensor must be calibrated before first use.

SE 707X/1-NMSN

14163 Berlin

00

44

BVS 10 ATEX E089  X

II 1G  Ex ia IIC T3/T4/T6

SE 707X/2-NMSN

14163 Berlin

00

44

BVS 10 ATEX E089  X

II 1G  Ex ia IIC T3/T4/T6

orange-red ring

Ex-Sensors

11. Disposal

Please  observe  the  applicable  local  or 

national regulations concerning the dis-

posal of “waste electrical and electronic 

equipment”.