Removing and Installing the Calibration Chamber
To separate the calibration chamber, you require ZU 0754 or ZU 0740 Service Set (see description
on page 36). We recommend the ZU 0746 and ZU 0747 mounting aids for proper mounting of
the gaskets and scraper rings (see description on page 37).
1) Remove the screws
(screwdriver TX25).
2) Position a plier and loosen the
thread of the split calibration
chamber using a face-pin spanner
wrench.
3) Completely screw off the split cali-
bration chamber. Now, the gaskets
are accessible and can be checked
and replaced if required. Use the
ZU 0746 and ZU 0747 mounting
aids for mounting the gaskets and
scraper rings.
How to handle the mounting
aids is described in the respective
instruction manual.
4) To re-install the split calibration
chamber, screw the parts together
using plier and face pin spanner
wrench and secure them with
screws.
Note:
The calibration chamber parts
must be firmly screwed together
(until the stop is reached) before
it can be secured with the two
screws.
Face-pin spanner wrench no. 5
included in accessory:
ZU 0740, ZU 0754
Mounting pliers
included in accessory:
ZU 0740, ZU 0754