![Knick SE102-AN Instructions For Use Download Page 2](http://html.mh-extra.com/html/knick/se102-an/se102-an_instructions-for-use_1980112002.webp)
1
Sensor head
2
Serial number
3
Oring seal for quiver
4
Filling hole with closure
5
Nameplate
6
Glass body
7
Junctions
8
pH glass
9
Watering cap, transport lock
10
"Watering cap,
for use with Portavo quiver"
a
6
SE102-AN
Manual
Instructions for Use of the SE102-AN pH Sensor
!
WARNING – Failure to observe this warning
may result in serious injury.
Read these instructions for use, observe the Specifi
cations, and follow the Safety Instructions.
1 Safety Instructions
Hazards due to temperature or aggressive media are
possible, depending on the location of use. There
fore, the operation and servicing of the sensor shall
only be carried out by suitably trained personnel
authorized by the operating company.
Be sure to wear safety goggles and gloves when
cleaning or calibrating a sensor using acid or alka
line solutions.
2 Sensor Description and Intended Use
The SE102AN pH sensor is equipped with 1 m fixed
cable. Connection to the transmitter (e.g. Portavo) is
made via a DIN 19262 coax connector and a 4 mm
spring connector.
The integrated temperature detector provides auto
matic temperature compensation during calibration
and operation of the sensor. The sensor is designed
for simultaneous measurement of pH and tempera
ture in industrial processes, particularly in the field of:
drinking water, wastewater, water processing.
The markings on the sensor or on the packaging label
include the following information:
Manufacturer of sensor
SE102AN
Model
0 … 14
pH range
5 … 100 °C
Temperature range
CE conformity mark
The quality certificate and the packaging label pro
vide the serial number and the sensor data of the final
inspection. The sensor fits into the sensor quiver of a
portable transmitter from the Portavo series.
The Oring
3
seals the quiver.
3 Commissioning
•
On unpacking, check the sensor for mechanical
damage. Report any damage to your Knick service
team.
•
Remove the watering cap and briefly rinse the sen
sor with pure water. After rinsing, the sensor should
only be dabbed dry. Do not rub the pHsensitive
glass, since this can lead to electrostatic charging
and sluggish response times.
•
Connect the sensor cable with the transmitter.
4 Operation
Note: Always open the closure
4
before calibrating
or measuring. Refill electrolyte if necessary.
4.1 Calibrating the Sensor
2point calibration is recommended for the SE102MS
sensor. The sensor is dipped successively into two
different buffer solutions with given pH values (e.g.,
pH 7.00 and pH 4.00). Please refer to the user manual
of the pH transmitter for further details.
4.2 Positioning
The sensor can be operated vertically or inclined up
to 45° off horizontal.
5 Maintenance and Cleaning
Never keep or store the sensor dry! You should store
the sensor with fitted watering cap containing elec
trolyte (3 mol/l KCl). If the sensor is to be inserted into
the quiver of a Portavo series measuring device, the
screwedon watering cap (9) must be replaced by the
included flexible watering cap (10).Close the filling
hole
4
when storing the sensor. Clean the sensor
(glass membrane and junction) with clean water after
each production cycle and regularly in case of heavy
pollution.
6 Specifications
Model:
SE102AN
pH:
0 … 14
Temperature:
5 … 100 °C
Length:
110 mm
Junction:
Ceramic
Reference system:
Ag/AgCI
with liquid electrolyte
Electrolyte, refillable: 3 mol/l KCl
7 Recommended pH Buffer and
Electrolyte Solutions
Order No.
CaliMat buffer
4.00 (20 °C)
CSP0400/250
CaliMat buffer
7.00 (20 °C)
CSP0700/250
Electrolyte
3 mol/l KCl
ZU 0960
For more buffer solutions and electrolytes, see
www.knick.de
8 Disposal
Observe the applicable local or national regulations
for disposal.
9 Connector plug
Ø 12 mm
110 mm
3
1
7
4
8
5
2
Read before commissioning.
Keep for future use.