Knick SE 715 Series Instructions For Use Download Page 2

SE 715/1-MS

Pmax: 3 bar tmax: 45 °C

SE 715/1-MS

Pmax: 3 bar tmax: 45 °C

SE 715/1-MS

Pmax: 3 bar tmax: 45 °C

SE 715/1-MS

Pmax: 3 bar tmax: 45 °C

SE 715-MS

Manual

Instructions for Use of the SE 715 Series Oxygen Sensors

1

Sensor connector (Memosens®)

2

19 mm A/F, serial number

3

PG 13.5 thread

4

Anode 

5

Cathode

6

Membrane cap

7

ZU0939 adapter

!

WARNING – Failure to observe this warn-

ing may result in serious injury.

The safety alert symbol on the nameplate 

means:

Read these instructions for use, observe the Spec-

ifications, and follow the Safety Instructions. 

The SE 715-MS sensor is equipped with Memosens 

technology. The benefits are a safe function due to 

perfect galvanic isolation, digital data transmission 

and an integrated sensor diagnostics, among others.

The connection is insensitive to dirt and humidity. 

The calibration data, operating time, sensor desig-

nation, and serial number are stored in the sensor 

head. The serial number is also printed in the quality 

certificate and on the packaging label.

1  Safety Instructions

The sensor contains liquid electrolyte. Be sure to 

wear safety goggles and gloves during mainte-

nance work, e.g., when replacing the membrane or 

the electrolyte.

The sensor is designed to withstand a maximum 

operating pressure of 3 bar and it meets the PED 

requirements (Pressure Equipment Directive 97/23/

EC, Article 3, Sec. 3). 

Please take extra care during installation to en-

sure that neither the polymer body nor the thread 

(PG 13.5) suffer any damage.

2  Intended Use

The sensor is designed for simultaneous measure-

ment of dissolved oxygen and temperature in indus-

trial processes, particularly in the field of: drinking 

water, wastewater, aeration, water processing, venti-

lation control, fish farming, aquariums.

3  Installation and Commissioning

 • On unpacking, check the sensor for mechanical 

damage. Report any damage to your Knick service 

team.

 • Remove the watering cap and briefly rinse the sen-

sor with clean water.

 • Install the sensor in the fitting as described in the 

user manual of the respective fitting or attach the 

ZU0939 adapter for storing the sensor in a sensor 

quiver of a portable meter from our Portavo series. 

Be careful not to damage the sensor membrane.

 • The sensor can be operated vertically or inclined up 

to 15° off horizontal.

 • Connect sensor and cable.

4  Operation

4.1 Calibrating the Sensor

A calibration in air is recommended for the SE  715 

sensor. To do so, place the sensor in air. Rinse the sen-

sor with clean water and carefully dab the membrane 

dry. Please refer to the user manual of the oxygen 

transmitter for further details. The resulting calibra-

tion data (slope, temperature, time of calibration) are 

saved directly in the sensor. Memosens sensors can 

be precalibrated in the lab. Therefore, calibration on 

site is not necessary. 

4.2 Temperature Detector 

The integrated temperature detector is intended for 

automatic compensation of the measurement signal 

and not for any high-precision and safe temperature 

indication or control of the process temperature. 

5  Maintenance and Cleaning

Observe the safety instructions given above when 

performing maintenance work on the sensor.

5.1  Replacing the Membrane

Hold the sensor vertically and carefully screw off the 

membrane cap. Fill a new membrane cap with elec-

trolyte (membrane caps and electrolyte ampoules in-

cluded in supply). Carefully screw the membrane cap 

on again. Rinse the sensor with clean water.

5.2 Cleaning the Cathode

Visible deposits on the cathode can be removed with 

fine sandpaper (included in supply). To do so, lay the 

sandpaper on a horizontal surface and place a drop of 

water on the paper. Remove the membrane cap from 

the sensor and move the sensor in a circular motion 

over the sandpaper. Hold the sensor vertically and 

exert no pressure. Rinse the sensor with clean water, 

fill a new membrane cap with electrolyte and careful-

ly screw the membrane cap on the sensor. Rinse the 

sensor with clean water.

5.3 Cleaning the Anode

The anode (silver/silver chloride) has a brown to violet 

color and does not require cleaning.

Clean the sensor with clean water after each produc-

tion cycle and regularly in case of heavy pollution.

6  Specifications

Model Code

The markings on each sensor or on the packaging 

label include the following information:

SE 715/ 1 - MS 

Model designation 

  

Sensor connector 

   

MS: Memosens®

  

Length 

   

1: approx. 120 mm

Further Data
Measuring range 

0 ... 20 mg/l

Resolution

20 µg/l

Temperature

-5 ... 45 °C

Pressure, relative

max. 3 bar

Membrane material silicone
Body material

polysulfone

Mounting

PG 13.5

Temp detector

NTC 30 kΩ

7  Package Contents

 • Sensor
 • Quality certificate
 • User manual
 • Maintenance kit, consisting of:

 

2 membrane caps

 

6 ampoules with electrolyte

 Sandpaper

 

Tissue paper

 • ZU0939 adapter for Portavo quiver

8  Disposal

Observe the applicable local or national regulations 

for disposal.

A

ppr

ox. 120 mm

1

6

3

4

Ø 12 mm

2

Knick Elektronische Messgeräte  

GmbH & Co. KG 

Beuckestraße 22

14163 Berlin

Germany

Phone:  

+49 30 80191-0

Fax:  

+49 30 80191-200

Email:  

[email protected]

Internet: www.knick.de

MEMO

SENS

5

7