Knick SE 558X/*-NMSN Series Manual Download Page 3

6

1

3

Ø 12 mm

120 / 225 mm

4

5

2

7

SE 558X/*-NMSN

SE 558X/*-NVPN

Manual

4.2 Temperature Detector 

The integrated temperature detector (see “Further 

Data”) is intended for automatic compensation of 

the pH signal and not for any high-precision and 

safe temperature indication or control of the process 

 temperature. 

5  Maintenance and Cleaning

Carefully rinse the pH glass (sensor tip) and junction 

with pure water after each operating cycle. Under no 

circumstances must measuring solution be allowed 

to dry on these parts! 

When the sensor is not in operation, store it in elec-

trolyte (3 mol/l KCl). If a sensor is stored dry for some 

time by mistake, let it soak in electrolyte for several 

hours and then test its function before the next use.

6  Specifications

Model Code

The markings on each sensor or on the packaging 

label include the following information:

SE 558X/ 1 - NMSN  Model designation 

  

 

Sensor connector 

  

 

VP: VarioPin

 

 

 

MS: Memosens®

  

 

Solution ground

 

 

 

N: without

  

 

Length 

 

 

 

1: 120 mm

 

 

 

2: 225 mm

  

 

Ex approval

 

 

 

X: Yes

Further Data

pH range

0 ... 14

Operating tem-

perature

-5 ... 100 °C

Pressure, relative

-1 ... 3 bar

Junction

Ceramic (3x)

Electrolyte

Gel with KCl reservoir

Reference system

Ag/AgCl

Sensor material

Alpha glass
medium impedance,  
universal glass,  
fluoride resistant

Body material

Glass

Mounting

PG 13.5

Temp detector

VP connector: Pt1000

MS connector: NTC 30 kΩ

7  Disposal

Observe the applicable local or national regulations 

for disposal.

091744

Instructions for Use for the SE 558 Series pH Sensors

WARNING – Failure to observe this warn-

ing may result in serious injury.

The warning symbol on the rating plate 

means:

Read these instructions for use, observe the Speci-

fications, and follow the Safety Instructions. 

1  Safety Instructions

1.1 All Applications – All Sensor Variants

Hazards due to pressure, temperature, aggres-

sive media or explosive atmosphere are possi-

ble, depending on the location of use. There-

fore, the installation, operation, and servicing of 

the sensor shall only be carried out by suitably 

trained personnel authorized by the operating 

company. 

1.2 Hazardous Areas – All Sensor Variants

Observe all applicable local codes and stan-

dards for the installation of electrical equipment 

in hazardous locations. For orientation, please 

refer to IEC 60079-14, EU directives 2014/34/EU 

and 1999/92/EC (ATEX), NFPA 70 (NEC), ANSI/

ISA-RP12.06.01. The electrical and thermal pa-

rameters of the sensors must be adhered to.

1.3 Hazardous Areas – Sensors with Memosens 

Connector

Memosens Ex sensors are marked by an orange- 

red ring. Combined with a model CA/MS-***X** 

or model CA/MS-***X**-L measuring cable or 

a certified measuring cable which is identical 

in hardware and function, the sensor may be 

connected to a suitable measuring device, as 

described in the Certificates BVS 15 ATEX E141 X 

and IECEx BVS 15.0114X.

1.4 Hazardous Areas – Sensors with VP 

 Connector

The sensors shall only be operated with an 

approved device in floating, intrinsically safe 

circuits. 

2  Intended Use

The low-maintenance sensors of the SE 558 series are 

particularly well suited for continuous measurement 

of pH in low-conductivity media such as boiler feed-

water, condensate, ultrapure water, WFI (water for 

injection), and cooling water. They are low-mainte-

nance pH sensors with gel reference electrolyte. The 

sensors are equipped with a temperature detector for 

automatic temperature compensation. More features 

for optimum functionality are 3 ceramic junctions, the 

gel reference electrolyte and a KCl salt reservoir that 

protects against leaching.

3  Installation and Commissioning

 •

On unpacking, check the sensor for mechanical 

damage. Report any damage to your Knick service 

team.

 •

Remove the watering cap and briefly rinse the 

sensor with clean water. After rinsing, the sensor 

should only be dabbed dry with a tissue. Do not 

rub the pH-sensitive glass, since this can lead to 

electrostatic charging and sluggish response times.

 •

Check the space behind the pH-sensitive glass for 

the presence of any air bubbles and remove them 

by gently shaking the sensor up and down.

 •

Install the sensor in the fitting as described in the 

user manual of the respective fitting.

 •

Connect sensor, cable, and transmitter.

4  Operation

4.1 Calibrating the Sensor

2-point pH calibration is recommended for the SE 558 

sensor. First remove the watering cap. Then dip the 

sensor successively into two different buffer solutions 

with given pH values (e.g., CaliMat pH 7.00 and pH 

4.00) and calibrate the pH transmitter to these buffer 

values. Please refer to the user manual of the pH trans-

mitter for further details.

8

9

TA-SE558-MSVP-KNX03  20161019

1

Sensor connector 

a: Memosens®

b: VarioPin (VP)

2

19 mm A/F, serial number

3

PG 13.5 thread

4

PVDF compression ring 

5

EPDM-FDA O-ring (11.5 x 2.6 mm)

6

Rating plate

7

KCl reservoir

8

Junction (3x)

9

pH glass

Knick Elektronische Messgeräte 

GmbH & Co. KG 

Beuckestraße 22

14163 Berlin

Germany

Phone: +49 30 80191-0

Fax: +49 30 80191-200

Email: [email protected]

Internet: www.knick-international.com

a

b

!

pH 0 to 14, -1 to 3 bar

, -5 to 100°C

SE 558/1

Summary of Contents for SE 558X/*-NMSN Series

Page 1: ...iche Sensoren mit Memosens Steckkopf Memosens Ex Sensoren sind durch einen orangeroten Ring gekennzeichnet Der Sensor in Verbindung mit dem Messkabel Typ CA MS X oder Typ CA MS X L oder ein in Hardware und Funktion identisches und be scheinigtes Messkabel darf an ein geeignetes Messgerät wie in den Zertifikaten BVS 15 ATEX E141 X und IECEx BVS 15 0114X beschrieben angeschlossen werden 1 4 Ex Berei...

Page 2: ...schlossenen Stromkreise und beträgt 427 K W in Luft Die maximal zulässige Umgebungstemperatur beträgt 40 C Temperaturklasse Maximale Eingangsleistung Pi Maximal zulässige Prozesstemperatur T6 50 mW 55 C T4 110 mW 80 C T3 200 mW 110 mW 100 C 135 C Für Sensoren ohne Temperaturfühler S8 Kopf Die im Hinblick auf die Temperaturklasse maximal zu berücksichtigende Erwärmung ist vernachläs sigbar klein Di...

Page 3: ...be adhered to 1 3 Hazardous Areas Sensors with Memosens Connector Memosens Ex sensors are marked by an orange red ring Combined with a model CA MS X or model CA MS X L measuring cable or a certified measuring cable which is identical in hardware and function the sensor may be connected to a suitable measuring device as described in the Certificates BVS 15 ATEX E141 X and IECEx BVS 15 0114X 1 4 Haz...

Page 4: ...rnally connected current circuits and is 427 K W in air The maximum permissible ambient temperature is 40 C Temperature class Maximum input power Pi Maximum permissible process temperature T6 50 mW 55 C T4 110 mW 80 C T3 200 mW 110 mW 100 C 135 C For Sensors without Temperature Detector S8 Connector The maximum heating to be considered with regard to the temperature class is negligibly low The max...

Reviews: