Uniclean 900(X) Parameter Setting
Functional Capabilities
Setting
Adjustable parameters
Page
•
Control
Off, Unical 9000, Uniclean 900
•
Time control
Fixed interval, Week program:
Fixed interval:
Select program, interval
Week program:
Configuration of program flows
•
Program flow
Individual adaptation of program steps for:
Rinsing, Cleaning, Parking, Service
•
Installation
Meas. procedures
Continuous / Short-time
External control (DCS)
Off, On
DCS inputs (36...39)
active 10...30 V or active < 2 V
M/S input (42/43)
active 10...30 V or active < 2 V
A/M input (40/41)
active 10...30 V or active < 2 V
Output DCS 34
Measuring, Alarm
DCS outputs (31…34)
N/O, N/C
•
Sensor detection
On, Off
•
Immersion lock
Off, Sensocheck Glass el
•
Access manual control
Access code required. All valves can be actuated separately.
•
Probe
Probe type (SensoGate, Ceramat, others)
Move time max.
Adjustable; default setting 0015 s
Sealing water
On, Off
Cavity rinsing
Off, Interval, Continuous
Check interval
On, Off (On: Entry: Check after x travels)
Maintenance interval
On, Off (On: Entry: Maintenance after x travels)
•
Media adapter (I ... III)
(up to 3 metering
pumps)
Each:
Pump (On, Off), Medium, Displaced volume, Residual volume
•
Add. media (1 ... 2)
Monitoring of medium (On, Off, Medium)
•
Start-up
Yes, No
•
System forecast
On, Off
Pg 100
Pg 108
Pg 113
Setting the Uniclean 900(X)
parameters
Pg 88
Summary of Contents for Protos 3400
Page 12: ...12 ...
Page 160: ...160 ...
Page 177: ...177 Appendix Buffer Table Mettler Toledo ...
Page 179: ...179 Appendix Buffer Table DIN 19267 ...
Page 181: ...181 Appendix Buffer Table Techn Buffers to NIST ...
Page 182: ...182 Appendix Buffer Table Hamilton ...
Page 183: ...183 Appendix Buffer Table Kraft ...
Page 184: ...184 Appendix Buffer Table Hamilton A ...
Page 185: ...185 Appendix Buffer Table Hamilton B ...
Page 186: ...186 Appendix Buffer Table HACH ...
Page 206: ...206 Index ...
Page 207: ...207 Index ...