74
Display text
Time [s]
01: Probe in SERVICE
02: Cleaning agent
0020 s
03: Rinse water ON
0060 s
04: Rinse water OFF
0002 s
05: Purge air ON
0010 s
06: Purge air OFF
0002 s
07: Cal buffer 1 (pH 4.01)
08: Wait position
> Position will be held until next command (e.g., DCS)
09: Rinse water ON
0010 s
10: Rinse water OFF
0002 s
11: Probe in MEASURE
0005 s
12: Program end
Parameter Setting: Program Flows
Parking
Program step
Probe: Measure
Probe: Service
Rinse water
Buffer 4.01
(Pump I)
Purge air
(AUX1 valve)
Cleaning agent
(Pump III)
HOLD
DCS ML3
(Program running)
01
02
03
04
05
06
07
08
09
10
11
Stroke 1 Stroke 2
Exposure time
Wait
[t]
s
Message (display)
W
ait p
osition
Parking is started via a DCS input signal at input BIN3
of the Unical 9000 probe controller.