70
Parameter Setting: Program Flows
Cleaning (Continuous), Media Monitoring Off
Display text
Time [s]
01: Probe in SERVICE
02: Cleaning agent
0020 s
03: Rinse water ON
0060 s
04: Rinse water OFF
0002 s
05: Probe in MEASURE
0005 s
06: Program end
Program step
Probe: Measure
Probe: Service
Rinse water
Cleaning agent
(Pump III)
HOLD
DCS ML3
(Program running)
01
02
03
04
05
06
Stroke 1
Stroke 2
Exposure time
[t]
s
Cleaning (continuous) can also be started via a DCS input signal at input BIN1
of the Unical 9000 probe controller.
Message (display)
Parameter setting: Program
flows
Cleaning program