Installation and connection
Diaphragm Vacuum Pump N 828 / N 838
26
Translation of original Operating and Installation Instructions, English, KNF 121249-121517 09/19
Pumps with brushless DC motor
1. Compare the supply data with the data on the motor plate. For
maximum operating current of the pump see pump’s type
plate.
The voltage must not vary by more than + 10% and - 10% from
that shown on the pump label.
2. Electrical connect of the pump according to Tab. 12, page 27
(UN 8__ K_DC-B) and Tab. 13, page 28 (UN 8__ K_.29DC-B).
Note the proper polarity (see marking on the motor). Incorrect
lead connection will damage electronics of brushless DC mo-
tors (type designation ending with B). The supply wires have
inverse-polarity protection in the motor board for this purpose,
while the control-voltage wires do not have this protection func-
tion.
Control voltage may only be applied if the motor controller is
supplied with operating voltage. Otherwise damages can occur
on the motor controller.