Troubleshooting
Vacuum Pump System SC/SEM/SH/SR950
54
Translation of original Operating Instructions, English, KNF 125549-125552 07/18
10. Troubleshooting
DANGER
Extreme danger from electrical shock!
Disconnect the pump power supply before working
on the pump.
Make sure the pump is de-energized and secure.
Vacuum pump system as a whole: see table 9.
Only SC 950:
Hand terminal: see table 10.
Only SC 950:
Software operation: see table 11.
Vacuum pump system as a whole
Problem
Cause
Solution
Vacuum pump system is
switched on, but the mains
switch does not light up.
Mains cable not plugged in.
Plug the mains cable for the vacuum
pump system into a properly
grounded power socket.
No current in electricity
network.
Check the electricity supply.
Vacuum pump system
fuses blown.
1. Find and eliminate the cause of the
overload.
2. Replace the mains fuse(s) of the
vacuum pump system:
Push the locking clip on the fuse
drawer down (Fig. 2/
17
, p. 13 or
14
13
, p. 19 and
12
, p. 21) and pull the drawer
out.
Replace the defective fuse(s)
(for fuse specifications see chapter
4.1, page 9f; for fuse order numbers
see chapter 11, p. 59).
Push the fuse drawer back in until it
clicks into position.
Required vacuum not
generated even though pump
is running.
Leaks in vacuum chamber. 1. Run a test by closing the gas inlet
19
16
, p. 17,
14
, p. 19 and Fig. 6/
13
, p. 21).
If the pump reaches the required
vacuum, this confirms the presence
of leaks in the vacuum chamber.
2. Eliminate leaks in the vacuum
chamber.
O-ring is not sitting
correctly in the slot on the
suction side of the
collection flask.
Adjust O-ring.
Replace if defective (for order
Summary of Contents for SC 950
Page 63: ......
Page 64: ...KNF worldwide Find your local KNF partner on www knf com...