Diaphragm Vacuum Pumps N 920__.18 and N 920 KT.29.18G
Technical Data
Translation of original Operating Instructions, English, KNF 121416-121419 07/18
9
N 920 KT.29.18G
Pneumatic performance
Max. permissible operating
pressure [bar g]
0.5
Ultimate vacuum [mbar abs.]
2.0
≤ 5 with open gas ballast
Delivery rate at atm. pressure
[l/min]*
max. 21
Pneumatic connections
Hose connection [mm]
ID 10
Ambient and media temperature
Permissible ambient
temperature
+ 10 °C to + 40 °C
Permissible media
temperature
+ 5 °C to + 40 °C
Other parameters
Weight [kg]
8.5
Dimensions : L x H x W [mm]
324 x 226 x 158
Maximum permissible ambient
relative humidity
80 % for temperatures up to
31 °C, decreasing linearly to
50 % at 40 °C
Max. altitude of site [m above
sea level]
2000
Electrical Data
Automatic mains power ad-
justment
100-240 V
50/60 Hz
Max. operating current [A]
1.4
Power consumption pump [W] 135
Maximum permitted mains
voltage fluctuations
+/- 10 %
Protection class motor
IP20
Tab. 5
* Liters in standard state (1,013 mbar)