Installation, mounting and connection
LABOPORT Pumps and Systems
32
Translation of original Operating Instructions, english, KNF 121211-121345 01/10
7.7.5. Retrofitting from System SC to System SCC
The second vacuum controller is delivered complete with the
vent valve installed on a new support (length: 410 mm). Install
the existing vacuum controller on the new support.
Pump removed from operation (see chapter 8.2, page 35)
Pump disconnected from mains
Qty Material
1
Allen key 4 mm
1
Allen key 5 mm
See chapter 7.7.4 for the required hoses
Tab. 21
Removing the existing vacuum controller
The following item numbers refer to fig. 2, page 8.
1. Disconnect the tubes from the pump vent valve (
5
) as well as
from the controller (
2
) vent valve (
3
).
2. Pull out all plugs from the electrical supply unit (
1
).
3. Remove the vacuum controller from the support (
16
) after
loosening the clamp assembly.
4. Remove the lateral support from the main support after loosen-
ing the clamp assembly.
5. Loosen the allen screw at the foot of the main support.
6. Slide the support out of the baseplate (
17
) mounting slot as-
sembly.
Mount new vacuum controller
The following item numbers refer to fig. 2, page 8.
1. Slide the removed vacuum controller (
2
) into the slot on the
new support (
16
) and fix it in place with the clamp connection.
2. Slide the support’s attachment part into the baseplate’s (
17
)
lateral upward-pointing slotted bar.
3. Tighten the allen screw at the foot of the support.
4. Mount the lateral support onto the main support.
5. Slide two additional vacuum valves(
6
) onto the lateral support
mounting slot and fix the device with help of the clamping as-
sembly.
6. Lay hoses for system (see fig. 19).
7. Wire and connect with each other the electrical supply unit and
the vacuum controller (see chapter 7.5, page 25).
Conditions
Tools and material
Fig. 19: Tubing system SCC