Safety
Rotary Evaporator RC 900
8
Translation of original Operating Instructions, English, KNF 126291-126293 11/14
When ventilating the rotary evaporator with air or inert gas, be sure
to prevent formation of reactive or explosive media. The maximum
permissible operating pressure of the chilled condenser is 0.1 bar
rel.
Ensure that the evaporation flask rotates throughout the entire
heating phase (even during submersion/lifting) in order to avoid
heating only one side of the evaporation flask or experiencing a
vaporisation delay.
Ensure that the evaporation flask is immersed or lifted only at a low
speed. Increase speed only to the extent that no heating medium is
ejected from the heating bath.
All replacement parts should be properly stored and disposed of in
accordance with the applicable environmental protection regula-
tions. Ensure adherence to the pertinent national and international
regulations. This especially applies to parts contaminated with toxic
substances.
The rotary evaporator complies with the safety stipulations of
Directive 2004/108/EC for electromagnetic compatibility, Directive
2006/42/EC for machines, Directive 2011/65/EU (RoHS2) and
Directive 1999/5/EC (R&TTE). The following harmonised standards
are fulfilled:
DIN EN 61010-1
DIN EN 61010-2-010
DIN EN 61326-1
The rotary evaporator complies with the following according to
IEC 664:
overvoltage category II
contamination level 2
All repairs to the rotary evaporator must be carried out by the
responsible KNF Customer Service team.
Housing parts with voltage-caring parts may be opened by trained
personnel only.
Only use genuine parts from KNF for servicing work.
Ventilating the
rotary evaporator
Vaporisation delay
Environmental protection
Standards
Customer service and
repairs
Summary of Contents for RC 900
Page 57: ......
Page 58: ......
Page 59: ......
Page 60: ...KNF worldwide Find your local KNF partner on www knf com ...