13
Packaging and Transport
KNAUER
Degasser 20DG User Manual V7632A, Version 1.1
Scope of Delivery
2-channel BlueShadow Online Degasser 20DG
Accessories
User manual DE/EN
Power supply 24 V/60 W
Accessory kit
Use original parts and
original accessories
• Only use original parts and accessories made by KNAUER or
a company authorized by KNAUER.
Checking the Scope of Delivery
1. Check whether the device and accessories are complete.
2. If a part is missing, inform the technical support of KNAUER.
Technical support hotline of KNAUER:
European hotline
Languages: German and English
Available by telephone: 8 am to 5 pm (CET)
Phone:+49-(0)30-809727-0
Fax:+49-(0)30-8015010
E-mail contact:
Space Requirements
Side clearance to other devices:
If there is a device on one side, min. clearance of 5 cm.
If there are devices on both sides, min. clearance of
10 cm.
At least 30 cm gap to the rear of the device.
Installation Site
Ambient conditions of
the installation site
Air humidity: Below 90 % (non-condensing)
Temperature range: 4–40 °C; 39.2–104 °F
Sunlight: When setting up the device at the installation loca-
tion, make sure that it is protected against direct sunlight.