AZURA® Detector RID 2.1L / RID 2.1L HighFlow Instructions V6750
Product information 33
Procedure
1.
Start the flush mode.
2.
Check if the LED blinks green.
3.
After a period of about 30 seconds, stop the flush mode.
4.
Wait for the end of program.
The detector is now ready to operate.
Start the measuring.
7.6
Activate standby
The flow is turned off.
Note:
Malfunctioning system after repeated standby possible. After
repeatedly using the standby, switch off the power switch and back on
again, to reset the data storage.
Procedure
Figure
1.
Press the standby switch
2
for 5 seconds.
2.
Wait if the LED
1
shows a
blue light.
2
1
If you were successful, the right LED lights up blue.
To end standby, press the switch again.
7.7
Switching off the detector
The detector is designed for the use of different solvents. In case the
detector has not been used for several weeks, solvent residues may cause
damage. If you plan to take the detector out of operation for a longer
period of time, fill the capillary with alcohol (e.g. isopropanol) before
switching it off.
The back must be accessible.
Detector has been switched on.
The flow cell has been flushed (see 7.5 on page 32).
Result
Next Steps
Prerequisites
Result
Next Steps
Prerequisites