AZURA® Detector RID 2.1L / RID 2.1L HighFlow Instructions V6750
32
Product information
Device defect
Intruding liquids can cause damage to the device.
Place solvent bottles next to the device or in a solvent tray.
Moisten the cleaning cloth only slightly.
Note:
State of safe operation is achieved even after the power connec-
tion has been interrupted because of power cut or emergency shutdown.
Procedure
Figure
1.
Plug in the power supply cable
2
into the socket.
2.
Push the main switch
1
to ON.
The right and middle LED light up green.
Flush the flow cell.
7.5
Flushing the flow cell
Flush the system (with the flush valve open) with water for 60 minutes
before starting the device test.
The flush mode activates the magnetic valve to flush the reference cell
with eluent. The flow path with active flush mode is described in a sepa-
rate section (see 1.5.3 on page 5).
The flush valve can be activated or deactivated directly via software;
alternatively a flush mode can be configured in the software, so that the
flush valve is activated and after a specific time span (30 s, 60 s, 120 s, 400
s) will be deactivated automatically. The flush mode can be cancelled any
time with the OFF command. If the flush valve is active, the left LED blinks
green.
The default values are listed in a separate section (see 7.2 on page 27).
The detector is supplied with eluent.
The detector has been switched on.
Software or Mobile Control
Note:
The flush mode starts via software of Mobile Control. Flush with
the eluent which you plan to use for the next measuring. If you change
the eluent, make sure that the used eluent is mixable with the one used
previously. Otherwise flush the flow cell with a solvent which is mixable
with both eluents.
1
2
Result
Next Steps
Prerequisites
Tools