AZURA® Detector RID 2.1L / RID 2.1L HighFlow Instructions V6750
Product information 27
Color
Operating condition
Operation
flashes green
equilibrated
Wait until the device
is ready.
green
Ready for operation
Right LED
(Operational status)
green
Switched on
blue
standby
Exit with the standby
switch.
To start the standby, keep the switch pressed for 5 seconds.
Note:
Malfunctioning system after repeated standby possible. After
repeatedly using the standby, switch off the power switch and back on
again, to reset the data storage.
7.2
Default settings
Using the Mobile Control, you can reset the detector to its default
settings.
Parameter
Setting
Network settings
LAN DHCP port 10001
Unit
µRIU
Date/time
Current date/time
Data rate
1 Hz
Signal
direct
Time constant
2 s
LED power factor
1
Temperature
35 °C for RID 2.1L
off for RID 2.1L HighFlow
Analog output scale
0.5 µRIU/mV for RID 2.1L
5 µRIU/mV for RID 2.1L
HighFlow
Analog output offset
100 mV
Analog output data rate
20 Hz
Analog output time constant
0.05 s
Leak sensor
ON, sensitivity = low
Extended measuring range
OFF
Flushing
OFF
Event impulse length
1000 ms
Autozero mode
OFF
Standby