P 2.1L user manual V6840, version 3.6
Troubleshooting
29
Cannot initialize
LAN
Check cables and connections in local area net-
work.
Cannot operate
with an empty link
Create a link.
Cannot purge
during the run!
End method and start purging.
Cannot read data
from FRAM
Switch the device off and on. Inform the techni-
cal support of the manufacturer in case the sys-
tem message repeats itself.
Cannot read RTC
Switch the device off and on. Inform the techni-
cal support of the manufacturer in case the sys-
tem message repeats itself.
Cannot start time
table
Edit the data on the device display or in the chro-
matography software.
Cannot use non-
existing compo-
nent!
Change the
setup
settings or change the gradi-
ent in the program or in
setup
.
Cannot write data
on FRAM
Switch the device off and on. Inform the techni-
cal support of the manufacturer in case the sys-
tem message repeats itself.
Component set-
tings not compati-
ble with gradient
setup!
Change the
setup
settings or change the gradi-
ent in the program or in
setup
.
E
Error input acti-
vated
Device error, change device settings.
G
GUI communica-
tion failed
Switch the device off and on. Inform the techni-
cal support of the manufacturer in case the sys-
tem message repeats itself.
I
Instrument in stand-
alone mode
Change the entry in the Setup menu.
Change the entry in the chromatography soft-
ware.
Instrument remote
controlled
This entry is not executable. Quit software.
Invalid index in time
table
Change the entry in the program line.
Invalid line number
Change the entry in the program line.
Invalid time in time
table
Correct the time entry.
L
Leak sensor not
present
Switch the device off and then on.
If the leak sensor is still not present, contact the
Technical Support of the manufacturer.
Leak was detected
Switch off the device. Remove the leak and start
the device afterwards.
System message
Solution