Calibration
Airsensor Supplement, V6879
Calibration
You must calibrate the airsensor before use or in case the signal is unsta-
ble. For that purpose, make sure that you can operate the amplifier.
NOTE
During calibration, the green LED STAB blinks rapidly. If
calibrating has been successful, the same LED blinks
slowly for approximately 3 s.
Integrating into PurityChrom®
You can program the software to either start or stop the system after an
air bubble has been detected.
NOTE
You have to choose the Gameport Input 1, 2, 3, or 4
which equals the airsensor number.
Process
Figure
1. Move the switch
3
to DLY.
2. Press the button -
2
.
The STAB signal disappears. Both lights
should be turned off.
3. Move the switch to AUT.
4. Press the
1
where the tubing is
filled with air.
5. Press the button - where the tubing is
filled with liquid.
1
2
3