Technical Parameters
GLP
Detailed report including lamp recognition,
ope rating hours, lamp operating hours, number
of lamp ignitions
Display
Mobile Control (optional)
Ambient conditions
Temperature range 4–40 °C, 39.2–104 °F
Humidity: below 90 %, non-condensing
General
Power supply
100–240 V, 50–60 Hz, 75 W
Dimensions
361 x 158 x 523 mm (H x W x D)
Weight
13.8 kg
Leak sensor
yes
12.2 DAD 2.1L
Detection
Detector type
Diode array detector
Number of diodes
256
Pixel pitch
2 nm/diode
Detection channels
8 (digital), 4 (analog)
Light source
Deuterium (D
2
) lamp with integrated GLP chip
Wavelength range
190–700 nm
Spectral bandwidth
< 8 nm at H
a
line (FWHM)
Note:
digital bandwidth 1–32 nm
Wavelength
accuracy
± 1 nm
Wavelength
verifica tion
Internal holmium filter and deuterium lines
Noise
± 5 µAU at 254 nm
Drift
400 µAU/h at 254 nm
Linearity
> 1.6 AU at 274 nm, typically 2.0 AU
Time constants
0.0 / 0.1 / 0.2 / 0.5 / 1.0 / 2.0 / 5.0 / 10.0 s
Integration time
Automatic (5–1000 ms)
61
AZURA® Detector DAD 6.1L, DAD 2.1L, MWD 2.1L Instructions, V6700