background image

Measured Quantities – Current, Temperature
Current 

current input option

„X/100mA“

„X/5A“

measuring range      

temperature drift

150V CAT IV

150V CAT IV

permanent overload

burden power ( impedance)

Current Unbalance

measuring range

0 ÷ 100 %

measuring uncertainty

± 1 % of rdg or ± 0.5

reference conditions
measuring range
measuring uncertainty

THDI 

measuring range

0 ÷ 200 %

measuring uncertainty

THDI <= 100 %: ± 0.6

THDI > 100 %: ± 0.6 % of rdg

Temperature (internal sensor, measured value affected by the instrument power dissipation)

measuring range

- 40 ÷ 80°C

measuring uncertainty

± 2ºC

spec-IT-en

I

NOM 

(I

B

) – rated (basic) current

0.1

 

A

AC

5

 

A

AC

0.00025 ÷ 0.15 A

AC

0.0125 ÷ 7.5 A

AC

intrinsic uncertainty  (t

A

=23 ±2 ºC)

+/- 0.05 % of rdg ± +/- 0.05 % of rng

 +/- 0.03 % of rdg ± +/- 0.01 % of rng / 10 ºC

measurement category

1 A

AC

10 A

AC

peak overload

1 second, maximum repetition 
frequency > 5 minutes

10 A

AC

90 A

AC

< 0.001 VA (Ri

 

< 0.1 Ω)

< 0.001 VA (Ri

 

< 0.1 Ω)

Harmonics & Interharmonics up to 50

th

 order (40

th

 order @ 60 Hz)

other harmonics up to 1000 % of class 3 acc. to IEC 61000–2-4 ed.2

500 % of class 3 acc. to IEC 61000–2-4 ed.2

Ih <= 10 % I

NOM

: ± 1 % I

NOM

Ih > 10 % I

NOM

: ± 1 % of rdg

12

Summary of Contents for SMC 133

Page 1: ...cription 1 2 Operating the Meter 2 2 1 Safety requirements when using SMC 133 2 2 2 Used symbols 2 2 3 Installation of the instrument 4 2 3 1 Supply voltage 4 2 3 2 Measured voltage 5 2 3 3 Measured c...

Page 2: ...nfrastructure and also remote load management Two different user interfaces are offered SMC 133 D with a small embedded color display on the front panel for simple local value reading and an unobtrusi...

Page 3: ...nd this person must know in detail the operation principles of the equipment listed in this description When the device is being connected to the parts which are under dangerous voltage it is necessar...

Page 4: ...quipment Earth ground To identify an earth ground terminal in cases where neither the symbol 5018 nor 5019 is explicitly required Class II equipment To identify equipment meeting the safety requiremen...

Page 5: ...d or the instrument might overheat and or the internal temperature value measured may be influenced A connection wire s maximum cross section area is 2 5 mm2 in case of all screw terminals The SMC 133...

Page 6: ...urrent transformers must be con nected to the terminal pairs I11 I12 I21 I22 I31 I32 with a cable of maximum cross section area 2 5 mm2 and a maximum length of 3 m X 100mA current input option The sup...

Page 7: ...surements star delta and single phase feeders Option E left for Ethernet port remote communication option U right for local USB communication port all instruments provide RS485 serial line Option S 10...

Page 8: ...ernal power supply can damage the instrument DI O as a digital input DI is connected to the selected I O terminal D1 or D2 and a common terminal C1 or C2 with external power supply in the following wa...

Page 9: ...to download The actual download will start by pressing the Download All button When finished the archive can be viewed in the ENVIS application User can open the downloaded file directly from ENVIS D...

Page 10: ...N A DC auxiliary voltage range power supply 3 5 W 7 VA 3W overvoltage category III 50 V II 100 V III 300 V II 600 V internal backup power supply adjustable 1 3600 s N A pollution degree 2 maximum ope...

Page 11: ...supply backup battery Li ion 18350 charging deactivated under 0 C protection class II weight max 0 25 kg spec Other en EN 61000 4 2 6 kV 8 kV EN 61000 4 3 10 V m up to 3 GHz EN 61000 4 4 4 kV EN 6100...

Page 12: ...certainty 0 3 of rdg or 0 3 THDU measuring range 0 20 measuring uncertainty 0 5 reference conditions measuring range measuring uncertainty twice the levels of class II acc to IEC 61000 4 7 ed 2 spec U...

Page 13: ...measured value affected by the instrument power dissipation measuring range 40 80 C measuring uncertainty 2 C spec IT en INOM IB rated basic current 0 1 AAC 5 AAC 0 00025 0 15 AAC 0 0125 7 5 AAC intr...

Page 14: ...UNOM x INOM reference conditions A ambient temperature tA U I for active power PF cos for reactive power 23 2 C U 80 120 UNOM I 1 120 INOM PF 1 00 PF 0 00 0 5 of rdg 0 005 PNOM PF cos uncertainty 0 0...

Page 15: ...al Inputs type Opticaly isolated unipolar Cx Dx maximum voltage voltage for logical 1 voltage for logical 0 input current spec IO en 35 VDC 30 mADC dynamic parameters pulse output pulse duration gap d...

Page 16: ...tems s r o Tr dr M Hora kove 559 460 05 Liberec 7 Czech Republic Tel 485 130 314 E mail kmb kmb cz Web www kmb cz The product must be in proper packaging to prevent damage during transit A description...

Reviews: