Adjustment Work
8.1
Safety Notice for Adjustment Work
WARNING
Risk of death by electrocution!
Power supply lines are installed in the suspension arm system. Touching live parts can cause
life-threatening electric shock.
Disconnect the unit from the mains before beginning any adjustment work:
•
For all types of installation or adjustment work, turn off the suspension arm system and
secure it against accidental reactivation.
•
Ensure that all the devices connected to the M6 service head are disconnected from the
power supply.
8.2
Adjusting the Mechanical Brake for the M6 service head
•
Observe the safety notice, see section 8.1 “Safety Notice for Adjustment Work”, page 86.
In all versions, the mechanical brake (friction brake) on the console tube holds the
M6 service head in the selected position.
•
Select the braking force of the M6 Service Head (with the device on the shelf) so that it
latches in every position and can be easily adjusted.