Kleinwächter PGT 51 Manual Download Page 6

 

Manual PGT 51 (PGT 52) 

Measure Operating 

 

 

short instruction 

 

 

 

 

At the bottom 

Limits 

(JUM1 OFF) 

 

The lower and upper limit will be 

Limits :  

 

lower limit

 

 

 

upper limit

 

 

 

additional limits 

Delivery parts 

 

PGT-51/2 include grounding cable with ring connector

Power plug 7,5V-9V= / 150-200mA

Manual 

 

Optional accessories 

   

RRP

 

 

 

 

lower and upper resistor. 

 

 

 

If the green LED shines ==> all is o.k.

If the red LED shines, check the correct position of your wrist strap

The measure starts automaticly after 10 seconds

Connect it to the popper on your wrist strap

Disconnect the wrist strap cable from Park position

 

At the bottom of the housing there is a short user manual !

The lower and upper limit will be calibrated by the RRP 50 box.. 

lower limit  < 750 kOhm 

upper limit   >  3,2 MOhm 

additional limits available ! 

clude grounding cable with ring connector 

200mA 

RRP-50 Reference resistor box with 

lower and upper resistor.  

(JUM1 OFF) 

 

If the green LED shines ==> all is o.k.

If the red LED shines, check the correct position of your wrist strap

The measure starts automaticly after 10 seconds

Connect it to the popper on your wrist strap

Disconnect the wrist strap cable from Park position

housing there is a short user manual !

 

Summary of Contents for PGT 51

Page 1: ...MANUAL Personal Grounding Permanent Personal Grounding Tester MANUAL Grounding Tester PGT 5 Permanent Personal Grounding Tester PGT 52 includes Relay Output 51 52...

Page 2: ...st be connected to the grounding point of the table or table mat Only use included power supply plug Limits can be changed by user very easy It can be done by calibrated resistors or by a reference wr...

Page 3: ...eans the unit recognizes whether the wrist strap is applied The result is shown by two LED s green LED flashes WST in Park position shines WST applied and into the limits red LED flashes WST resistor...

Page 4: ...ications Dimension L x B x H 96mm x 60mm x 23mm L x B x H Weight app 150 g Power Supply Power plug 7 5V 9V 150 200mA Display 1 LED green 1 LED red 1 buzzer Relay Output PGT 52 Potential free Relay Con...

Page 5: ...he select limits The instrument really measures the capacitive resistor but is calibrated by a ohmic resistor the actually limit can t be definite exactly But we have a reproducible measure Is the mea...

Page 6: ...t it to the popper on your wrist strap Disconnect the wrist strap cable from Park position At the bottom of the housing there is a short user manual The lower and upper limit will be calibrated by the...

Page 7: ...ld with a Pin the hidden switch H connect the power plug and release switch H LED o k green on RRP 50 lower UGW connect to ground No WST connect Measure input free Presses briefly switch H LED o k gre...

Page 8: ...only simulate the capacitive attenuation of a person wearing the wrist strap Suggested calibration cycles annual Guarantee Items We provide a 24 month guarantee in case of proper application accordin...