OPERATING INSTRUCTIONS

CONNECTING TEST LEADS

Do not test if leads are improperly seated. Results could cause 
intermittent display readings. To ensure proper connection, firmly 
press leads into the input jack completely. Lead guard should be 
flush with the meter’s faceplate.

TESTING IN CAT III / CAT IV MEASUREMENT LOCATIONS

Ensure the test lead shield is pressed firmly in place. Failure to use 
the CATIII / CATIV shield increases arc-flash risk.

TESTING IN CAT II MEASUREMENT LOCATIONS

CAT III / CAT IV shields may be removed for CAT II locations. This 
will allow testing on recessed conductors such as standard wall 
outlets. Take care not to lose the shields.

PRESS

NO GAP

5/32" 

(4 mm)

.7" (18 mm)

CL210-1390114ART.indd   7

10/1/2015   3:42:25 PM

Summary of Contents for CL210

Page 1: ...Dwg Ref 1290186 Color Reference N A INSTRUCTION MANUAL ENGLISH CL210 FRAN AIS pg 29 ESPA OL pg 15 400A AC Auto Ranging Digital Clamp Meter AUTO RANGING DATA HOLD RANGE HOLD TEMPERATURE AUDIBLE CONTIN...

Page 2: ...n ambient working temp is outside of Accuracy Temp range Dimensions 8 66 x 3 03 x 1 61 220 x 77 x 41 mm Weight 9 88 oz 280 g including batteries Calibration Accurate for one year Standards Conforms to...

Page 3: ...gits 20 00A 10mA 2 0 10 digits 200 0A 100mA 400A 1A Frequency Range 50 to 60Hz Resistance 200 0 0 1 1 2 5 digits 2 000K 1 1 2 3 digits 20 00k 10 200 0k 100 2 000M 1k 20 00M 10k 2 0 5 digits Maximum In...

Page 4: ...tion when working with voltages above 25V AC RMS or 60V DC Such voltages pose a shock hazard To avoid false readings that can lead to electrical shock replace batteries when a low battery indicator ap...

Page 5: ...7 4 1 2000 count LCD display 7 RANGE button 2 Function selector switch 8 MAX Maximum button 3 Clamp 9 Data Hold button 4 COM jack 10 Clamp trigger press to open clamp 5 V jack 11 Arrow markings 6 Bac...

Page 6: ...The backlight does not automatically power OFF RANGE The meter defaults to auto ranging mode This mode automatically determines the most appropriate measurement range for the testing that is being co...

Page 7: ...etely Lead guard should be flush with the meter s faceplate TESTING IN CAT III CAT IV MEASUREMENT LOCATIONS Ensure the test lead shield is pressed firmly in place Failure to use the CATIII CATIV shiel...

Page 8: ...care should be taken to ensure that the clamp is completely closed with trigger 10 fully released and that the wire passes perpendicularly through the center of the clamp in line with the arrow markin...

Page 9: ...sure voltage The meter will auto range to display the measurement in the most appropriate range NOTE If appears on the LCD the test leads are being applied to the circuit in reverse Swap the position...

Page 10: ...to display the measurement in the most appropriate range NOTE When in a Resistance setting and the test leads are open not connected across a resistor or when a failed resistor is under test the disp...

Page 11: ...ple tip and the object being measured When thermocouple tip and object are in thermal equilibrium the measurement on the display will stabilize The meter will auto range to display the measurement in...

Page 12: ...ns before using MAINTENANCE BATTERY REPLACEMENT When indicator is displayed on LCD batteries must be replaced 1 Remove screw from battery door 2 Replace 3 x AAA batteries note proper polarity 3 Replac...

Page 13: ...ing the limits mentioned in the General Specifications section allow the meter to return to normal operating conditions before using WARRANTY www kleintools com warranty DISPOSAL RECYCLE Do not place...

Reviews: