12 Command list
lx
The maximum value memory is erased,
when the erasure time to external
Write:
xxlx (Ausgang 0 . . . 20 mA)
Response: „ok” when deleted,
otherwise
„no”
lz
Erase time of the maximum value memory
[s]:
0 = aus
1 = 0.01
2 = 0.05
3 = 0.25
4 = 1
5 = 5
6 = 25
7 = extern
8 = autom.
Command value: 1 digit
Read:
xxlz
Response: 1 digit
Write:
xxlz2 (erase time = 0.05 s)
Response: „ok” when accepted, otherwise
„no”
Limits:
xxlz?
Response: 1 digit min. 1 digit max. („08”)
m1
Adjust sub-measuring range
Command value:
4 digit Hex TMB start
4 digit Hex TMB end
Write:
xxm1003200FA
(TMB:
-50
. . . 250 °C)
Response: „ok” when accepted, otherwise
„no”
Limits:
xxm1?
Response: 4 digit min. 4 digit max., (corre-
sponds to measuring range)
Note:
The beginning and end must have a mini-
mum separation of 51 °C (dependent on the
adjusted measuring unit, see fh).
mb
Inquiry measuring range
Read:
xxmb
Response: 4 digit start, 4 digit end in hex,
dependent of adjusted unit (see
fh)
me
Inquiry sub-measuring range
Read:
xxme
Response: 4 digit start, 4 digit end in hex,
dependent of adjusted unit (see
fh) )
me
Inquiry measured value
Read:
xxmsaaa
Response: 5 digit, 10 x measuring value
without decimal point (03456 =
345.6 °C)
first position „-” with negative values, aaa =
Number with multiple queries, corresponds
to aaa times xxms, dependent on adjusted
measuring unit (see fh)
na
Inquiry device name
Read:
xxna
Response: KPE 660/5
pa
Inquiry block parameter
Read:
xxpa
Response: 11 digit
2 digit emissivity in % (rounded)
1 digit response time
1 digit erase time
1 digit current output configuration
2 digit nternal temperature (always in °C)
2 digit device address
1 digit baud rate
1 digit keyboard lock
KPE 660/5
35
Summary of Contents for KPE 660/5
Page 23: ...7 Operation of the pyrometer Figure 11 Menu plan configuration level Config KPE 660 5 19 ...
Page 25: ...7 Operation of the pyrometer Figure 14 Menu plan configuration level Switch 1 KPE 660 5 21 ...
Page 26: ...7 Operation of the pyrometer Figure 15 Menu plan configuration level Switch 2 22 KPE 660 5 ...
Page 42: ...12 Command list 38 KPE 660 5 ...
Page 63: ...7 Bedienen des Pyrometers Abbildung 11 Menüplan Konfigurationsebene config KPE 660 5 19 ...
Page 65: ...7 Bedienen des Pyrometers Abbildung 14 Menüplan Konfigurationsebene Switch 1 KPE 660 5 21 ...
Page 66: ...7 Bedienen des Pyrometers Abbildung 15 Menüplan Konfigurationsebene Switch 2 22 KPE 660 5 ...