background image

Hand-held Controller Operation

Turn on the power switch located at the  

 

foot of the machine.

Range of Motion Set

Hold down “EXT” or “FLEX” and press “UP” 

or “DOWN” arrow to set the limits. Full range 

is from –10° extension to 120° flexion.

Speed Set

Hold down “SPEED” and press “UP” or 

“DOWN” arrow to set speed. Speed can be 

adjusted in 10 degree increments ranging 

from 30° per minute to 150° per minute.

Pause Set

Hold down “DELAY” and press “UP” 

or “DOWN” arrow to set the length of 

time the unit will pause at the flexion 

and extension limits. Delay range is 

from 0 seconds to 30 seconds.

Press “START/STOP” to begin motion.

Press “START/STOP” again to stop motion.

Setting Lock

Press both “UP” and “DOWN” arrows at the 

same time to lock the setting, and again to unlock. 

When locked, a dot will show in the lower right 

corner of the display.

NOTE:  Locking mode retained 

even when power is turned off.

P/N 1.07080 Revision G                                                                                7 of 8                                                                                                    

Cleaning and Maintenance

With unit unplugged, clean housing and leg carriage with a mild soap solution and dry thoroughly. refer to the service manual for recommended maintenance.
caution:  avoid getting water inside the housing as it may damage the electrical components.

Summary of Contents for FLEX-MATE K500

Page 1: ...USER S MANUAL P N 1 07080 Revision G 1 of 8 4038 Weaver Court East Hilliard OH 43026 Ph 614 921 8866 Fx 614 921 8877 Toll Free 866 532 3534...

Page 2: ...Federal law restricts this device to sale by or on the order of a licensed health care practitioner CAUTION Work in a range of motion as instructed by your healthcare practitioner CAUTION If you expe...

Page 3: ...lug the transformer from wall Use only the external power supply provided with the FLEX MATE K500 in order to meet the requirement of UL 60601 1 Package Contents 1 FLEX MATE K500 1 Users Manual 1 Serv...

Page 4: ...cure the patient pads to the machine The Foot Pad has a pocket that slips over the Foot Plate The Tibia and Femur Padshavestrapsthatsecureto the black velcro pads 4 Measure the patient s involved leg...

Page 5: ...ee in the proper alignment 5 Loosen the adjustment knobs on the tibia bars and pull the foot plate out to a length greater than the patient s lower leg 6 Use the hand held controller to bring the mach...

Page 6: ...ate 2 Loosen the adjustment knobs on each side of the foot plate 3 Rotate the foot plate bracket inward 180 as shown 4 Reattach the foot plate to the foot plate bracket Adjust for plantar dorsiflexion...

Page 7: ...use at the flexion and extension limits Delay range is from 0 seconds to 30 seconds Press START STOP to begin motion Press START STOP again to stop motion Setting Lock Press both UP and DOWN arrows at...

Page 8: ...4038 Weaver Court East Hilliard OH 43026 Ph 614 921 8866 Fx 614 921 8877 Toll Free 866 532 3534...