Stylus Change Procedure - Stylus Removal and Replacement
KLA-Tencor P-16+ / P-6 User’s Guide
13-8
KLA-Tencor Confidential
0142530-000 AB
3/13/09
3.
While supporting the stylus and the stylus arm, gently tighten the clamp screw
to hold the stylus in place. Do not over tighten or damage can occur to the stylus
arm pivot. (See
4.
Remove the wrench from the clamp screw and replace it in its mount. Tighten
the thumbscrew to hold the wrench in place.
Figure 13.10
Supporting Stylus and Mount During Tightening Procedure
Figure 13.11
Sensor Assembly - Seating the New Stylus
these two points. See
Gently tighten the stylus clamp
screw to hold the stylus in place.
CAUTION
: Do not turn
this screw; it can result in
damage to the stylus.