Calibrations - P-16 Series Level Calibrations
The Level Calibration should be performed whenever one of the following conditions
are present.
System does not complete the initialization procedure
Changing sample locators (e.g.: Adding a precision stress locator)
Scans have excessiveraw non-level
Level Calibration Procedure
12
1.
From from any top level screen, click on the
Calibrations
icon
.
2.
Click
Level…
to open the Level Calibration screen.
Figure 12.13
Calibration Screen
Click
Level…
to open
the calibration screen.