KLA-Tencor P-16+ / P-6 User’s Guide
Calibrations - Step Height Calibration
11.
From the
Calibrations
screen, choose
Step Height…
(See
Step
Height Calibration Options
dialog box appears in the center of the window. (See
12.
Range
: Choose the range to be calibrated. Select the appropriate step height
standard for use in calibrating the selected range. (See the circled area in
Figure 12.11.)
13.
Multi-Scan Average
: This determines the number of times the profiler scans the
same feature during each scan procedure. Data from all scans are automatically
averaged and their average is presented as the scan result. Click the down-arrow
next to the
Multi-Scan Average
value box to display the menu.
Select the number of scans per calibration from the drop-down menu. (See
below.) The value should be at least 3, with 5 being optimum.
Figure 12.10
Calibration Menu in the Calibration Screen
Calibration screen,
choose
Step Height…
to
open the Step Height
calibration screen.