H o w t o U s e
K F 0 1
Series
Instruction Manual
2.3.12
[
F-92
]
External Input Selection
Accumulated flow external reset
:
The accumulated flow value will
reset to "0" when an external input signal is applied.
Auto-shift
:
The instantaneous flow rate will regard as the standard
when the external input signal is applied. The switch output
function operates relative to its change.
Auto-shift zero
:
The instantaneous flow rate is reset to zero to regard
as standard when the external input signal is applied.
The switch output function operates relative to its
change.
This function is only for output 1 corresponding to flow sensor action point.
When external signal is input, please connect the input wire to GND for
30 ms or more.
Press or button at Function Selection Mode to display
[
F-92
] [
inP
]。
Press button
External Input Selection
Press or button to select
external input function.
※
NOTE
:
1. This function is not available with Output Specification -02 and -04.
2. When external signal is input, the instantaneous flow
rate value will be shown ”
oooo
”.
Accumulated flow external reset
Auto-shift
Auto-shift zero
External Input
Selected Mode
L/min
L/min
Flashing
alternately
L/min
L/min
Flashing
alternately
L/min
L/min
Flashing
alternately
L/min
L/min
Flashing
alternately
L/min
L/min
Flashing
alternately
Press button to return to
Function Selection Mode
L/min