CHAPTER 2. PRE-CHECK PROCEDURES
INTRODUCTION
This chapter describes the pre-check procedures
for Microcell™ sets. Verifying the application and
checking the Microcell™ sets before installation
will ensure installation of properly working
equipment that will provide accurate monitoring
of vessel contents.
APPLICATION VERIFICATION
Prior to ordering Microcell™ sets, be sure to have
read the Microcell™ Selection Guide
(Kistler-Morse
®
#97-5023) and completed the
appropriate Application Data Form (Kistler-Morse
®
#97-5025 for Microcells™ on vertical column legs
or Kistler-Morse
®
#97-5024 for Microcells™ on
horizontal beams). A copy of the completed form
was returned with both the order acknowledgment
and equipment shipment. If you cannot locate the
form, contact Kistler-Morse
®
to get another copy
before proceeding. Review the information on the
form now to verify the application details.
Note
If the calculated stress on the Application Data
Form is outside the following ranges, this is a
special application:
3” Microcell™:
2,500 psi - 7,500 psi
(1.8kg/mm
2
- 5.3kg/mm
2
)
2” Microcell™: 3,750 psi - 11,250 psi
(2.6kg/mm
2
- 7.9kg/mm
2
)
Contact Kistler-Morse
®
before proceeding further
with a special application.
ORDER VERIFICATION
Prior to beginning installation, verify the order is
complete and assemble additional equipment
needed for the installation.
MICROCELL™ ORDER
The following are included with the order
(quantities dependent on application):
STANDARD
Microcell™ set, each complete with:
Sensor
Environmental Cover
#8-32 socket head cap screws (2)
#8 hardened flat washers (2)
JB1 or JB2 Junction Boxes, each complete with:
Terminal board
Watertight fittings (4)
Watertight plugs (for any cable openings that
will not be used)
Installation Kit, each complete with:
Microcell™ drill template with #8-32 socket
head cap screw
#29 drill bit
#8-32, 2-flute, spiral-point tap
Sikaflex 1A polyurethane sealant or Dow
Corning RTV 738 or RTV 739 and Material
Safety Data Sheet (MSDS)
Rust-inhibiting silicone grease
OPTIONAL
Insulation and insulation hardware (if best
performance is required for an outdoor
installation on column legs)
If any items are missing from the order, contact
Kistler-Morse
®
before proceeding. Substituting
parts without Kistler-Morse
®
approval may cause
system problems and will void the warranty.
Note
A signal processor and its manual are required
to calibrate the system. If using an existing
signal processor, this will not be included in
the order.
2