INSTALLATION AND MANTENANCE
Instruction manual PAR-LITE
22
4.3
Measuring modulated light
In the same way it is possible to shunt the PAR
LITE
with a
capacitor. This can be necessary when modulated light must
be measured. The PAR
LITE
has a response time much
smaller than 0,1s and much read-out equipment cannot
handle the resulting voltage pulses. At principle you must bring
the RC time above the period time of the lightpulses to get a
smooth signal. Due to the relative low impedance of the PAR
LITE
this need big capacitors. which can be problematic
because electrolytic capacitors are often useless due to their
battery effect. So use polyester, ceramic, etc types.